Automated 3-dimensional elastic registration of whole-body PET and CT from separate or combined scanners

Raj Shekhar, Vivek Walimbe, Shanker Raja, Vladimir Zagrodsky, Mangesh Kanvinde, Guiyun Wu, Bohdan Bybel
Journal of Nuclear Medicine 2005, 46 (9): 1488-96
16157532

UNLABELLED: Registration and fusion of whole-body functional PET and anatomic CT is significant for accurate differentiation of viable tumors from benign masses, radiotherapy planning and monitoring treatment response, and cancer staging. Whole-body PET and CT acquired on separate scanners are misregistered because of differences in patient positions and orientations, couch shapes, and breathing protocols. Although a combined PET/CT scanner removes many of these misalignments, breathing-related nonrigid mismatches still persist.

METHODS: We have developed a new, fully automated normalized mutual information-based 3-dimensional elastic image registration technique that can accurately align whole-body PET and CT images acquired on stand-alone scanners as well as a combined PET/CT scanner. The algorithm morphs the PET image to align spatially with the CT image by generating an elastic transformation field by interpolating quaternions and translations from multiple 6-parameter rigid-body registrations, each obtained for hierarchically subdivided image subvolumes. Fifteen whole-body (spanning thorax and abdomen) PET/CT image pairs acquired separately and 5 image pairs acquired on a combined scanner were registered. The cases were selected on the basis of the availability of both CT and PET images, without any other screening criteria, such as a specific clinical condition or prognosis. A rigorous quantitative validation was performed by evaluating algorithm performance in the context of variability among 3 clinical experts in the identification of up to 32 homologous anatomic landmarks.

RESULTS: The average execution time was 75 and 45 min for images acquired using separate scanners and combined scanner, respectively. Visual inspection indicated improved matching of homologous structures in all cases. The mean registration accuracy (5.5 and 5.9 mm for images from separate scanners and combined scanner, respectively) was found comparable to the mean interexpert difference in landmark identification (5.6 +/- 2.4 and 6.6 +/- 3.4 mm, respectively). The variability in landmark identification did not show statistically significant changes on replacing any expert by the algorithm.

CONCLUSION: We have presented a new and automated elastic registration algorithm to correct for nonrigid misalignments in whole-body PET/CT images as well as improve the "mechanical" registration of a combined PET/CT scanner. The algorithm performance was on par with the average opinion of 3 experts.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
16157532
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"