Cyclic loading comparison between biodegradable interference screw fixation and biodegradable double cross-pin fixation of human bone-patellar tendon-bone grafts.
Arthroscopy 2005 August
PURPOSE: The aim of this study was to compare ultimate load, yield load, stiffness, and displacement after cyclic loading of a cross-pin technique and an interference screw technique for the fixation of bone-patellar tendon-bone (BPTB) grafts in anterior cruciate ligament (ACL) reconstruction.
TYPE OF STUDY: Biomechanical in vitro study.
METHODS: The devices tested were 2 2.7-mm biodegradable pins (RigidFix; Ethicon, Mitek Division, Norderstedt, Germany) and biodegradable interference screws (Absolute; Innovasive Devices, Marlborough, MA). Each device was used for the fixation of 10 8-mm, 9-mm, or 10-mm sized human BPTB grafts in tunnels drilled in bovine knees. Ultimate load, yield load, stiffness, and displacement after cyclic loading (1,000 cycles between 50 and 250 N) were then evaluated.
RESULTS: All 8-mm grafts that were fixed with cross-pins failed after a mean of 124 cycles of load. The 9-mm and 10-mm grafts survived the cyclic loading protocol. Yield load and maximum load of the 10-mm groups (cross-pin and interference screw) were significantly higher than that of the 9-mm groups. There was no significant difference in maximum load, yield load, and stiffness between the cross-pin and interference screw fixation technique for 1 graft size.
CONCLUSIONS: The biomechanical data suggest that femoral fixation of 9-mm and 10-mm BPTB grafts using 2.7-mm biodegradable cross-pins leads to primary stability that is comparable to that of biodegradable interference screws. Fixation of 8-mm BPTB grafts using 2.7-mm biodegradable cross-pins had poor results.
CLINICAL RELEVANCE: The diameter of the bone block is the limiting factor for the final fixation strength and the cyclical survival when using cross-pins. We strongly recommend not using this technique for bone blocks smaller than 9 mm in diameter.
TYPE OF STUDY: Biomechanical in vitro study.
METHODS: The devices tested were 2 2.7-mm biodegradable pins (RigidFix; Ethicon, Mitek Division, Norderstedt, Germany) and biodegradable interference screws (Absolute; Innovasive Devices, Marlborough, MA). Each device was used for the fixation of 10 8-mm, 9-mm, or 10-mm sized human BPTB grafts in tunnels drilled in bovine knees. Ultimate load, yield load, stiffness, and displacement after cyclic loading (1,000 cycles between 50 and 250 N) were then evaluated.
RESULTS: All 8-mm grafts that were fixed with cross-pins failed after a mean of 124 cycles of load. The 9-mm and 10-mm grafts survived the cyclic loading protocol. Yield load and maximum load of the 10-mm groups (cross-pin and interference screw) were significantly higher than that of the 9-mm groups. There was no significant difference in maximum load, yield load, and stiffness between the cross-pin and interference screw fixation technique for 1 graft size.
CONCLUSIONS: The biomechanical data suggest that femoral fixation of 9-mm and 10-mm BPTB grafts using 2.7-mm biodegradable cross-pins leads to primary stability that is comparable to that of biodegradable interference screws. Fixation of 8-mm BPTB grafts using 2.7-mm biodegradable cross-pins had poor results.
CLINICAL RELEVANCE: The diameter of the bone block is the limiting factor for the final fixation strength and the cyclical survival when using cross-pins. We strongly recommend not using this technique for bone blocks smaller than 9 mm in diameter.
Full text links
Related Resources
Trending Papers
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app