Effects of cardiac resynchronization therapy on ventricular repolarization in patients with congestive heart failure

Thomas Berger, Friedrich Hanser, Florian Hintringer, Gerhard Poelzl, Gerald Fischer, Robert Modre, Bernhard Tilg, Otmar Pachinger, Franz X Roithinger
Journal of Cardiovascular Electrophysiology 2005, 16 (6): 611-7
15946359

INTRODUCTION: Biventricular pacing has been shown to improve the clinical status of patients with congestive heart failure, but little is known about its influence on ventricular repolarization. The aim of our study was to evaluate the effect of biventricular pacing on ECG markers of ventricular repolarization in patients with congestive heart failure.

METHODS AND RESULTS: Twenty-five patients with congestive heart failure, sinus rhythm (SR), and complete LBBB (6 females; age 61 +/- 8 years; NYHA class II-III; echocardiographic ejection fraction 21 +/- 5%; QRS > or = 130 ms) underwent permanent biventricular DDDR pacemaker implantation. A high-resolution 65-lead body-surface ECG recording was performed at baseline and during right-, left-, and biventricular pacing, and the total 65-lead root mean square curve of the QRST complex and the interlead QT dispersion were assessed. The QRS duration was increased during right (RV)- and left ventricular (LV) pacing (127 +/- 26% and 117 +/- 40%; P < 0.05), as compared to SR (100%) and biventricular pacing (93 +/- 16%; ns). The QTc interval was increased during RV and LV pacing (112 +/- 12% and 114 +/- 14%; P < 0.05) as compared to SR (100%) or biventricular pacing (99 +/- 12%). There was no effect on JT interval during all pacing modes. The T(peak-end) interval was increased during right (120 +/- 34%; P < 0.01) and LV pacing (113 +/- 29%; P < 0.05) but decreased during biventricular pacing (81 +/- 19%; P < 0.01). A similar effect was found for the T(peak-end) integral and the T(peak) amplitude. QT dispersion was increased during right ventricular (129 +/- 16 ms; P < 0.05) and decreased during biventricular pacing (90 +/- 12 ms; P < 0.01), as compared to SR (114 +/- 22 ms).

CONCLUSIONS: Using a high-resolution surface ECG, biventricular pacing resulted in a significant reduction of ECG markers of ventricular dispersion of repolarization.

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
15946359
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"