[Contribution of early systole to total antegrade flow volume in breath-hold phase-contrast flow measurements]

R P Kunz, F Oellig, F Krummenauer, K Oberholzer, D Neeb, T W Vomweg, G Horstick, K-F Kreitner
RöFo: Fortschritte Auf Dem Gebiete der Röntgenstrahlen und der Nuklearmedizin 2005, 177 (5): 637-45
15871078

PURPOSE: To evaluate the contribution of early systole for the assessment of antegrade aortic flow volume by breath-hold velocity-encoded magnetic resonance (MR) flow measurements.

MATERIALS AND METHODS: Expiratory breath-hold fast low-angle shot (FLASH) phase-contrast flow measurements (temporal resolution tRes 61 msec, shared phases) perpendicular to the proximal ascending aorta and short axis true fast imaging with steady-state precession (TrueFISP) cine MR ventriculometry (tRes 34.5 msec) were performed in ten healthy male volunteers on a 1.5 T MR system (Sonata, Siemens Medical Solutions). Antegrade aortic flow volume (AFV) and left ventricular stroke volume (LV-SV) were evaluated using Argus Ventricular Function and Argus Flow Software, version MR 2002B (Siemens Medical Solutions). A beta release of Argus Flow MR 2004A allowed interpolation of the flow up-slope during early systole to the preceding R-wave trigger. The respective intraindividual median differences between the AFV of each flow evaluation and LV-SV as well as between both AFV measurements were calculated and compared using the sign test for paired samples.

RESULTS: Non-interpolated AFV significantly deviated from LV-SV (p = 0.006), underestimating the latter by 13.1 mL (13 %). Interpolating aortic flow during early systole significantly increased AFV by 10.8 mL (13 %) compared to the flow evaluation which did not take early systole into account (p = 0.006). AFV with interpolation of early systolic flow agreed well with LV-SV (median difference - 3.0 mL or - 3 %, respectively), and no significant difference between these measurements was found (p = 1.0).

CONCLUSION: Flow during early systole contributes substantially to total forward flow volume in the ascending aorta. Interpolation of the early systolic up-slope is therefore recommended for the evaluation of breath-hold phase-contrast flow measurements.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
15871078
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"