A positron emission tomography/computed tomography (PET/CT) acquisition protocol for CT radiation dose optimization.
Nuclear Medicine Communications 2005 April
BACKGROUND: In current combined positron emission tomography/computed tomography (PET/CT) systems, high-quality CT images not only increase diagnostic value by providing anatomic delineation of hyper- and hypometabolic tissues, but also shorten the acquisition time for attenuation correction compared with standard PET imaging. However, this technique potentially introduces more radiation burden to patients as a result of the higher radiation exposure from CT.
METHODS: In this study, the radiation doses delivered from typical germanium-based and CT-based transmission scans were measured and compared using an anthropomorphic Rando Alderson phantom with insertions of thermoluminescent dosimeters. Image geometric distortion and quantified uptake values in PET images with different manipulating CT acquisition protocols for attenuation correction were also evaluated.
RESULTS: It was found that radiation doses during germanium-based transmission scans were almost negligible, while doses from CT-based transmission scans were significantly higher. Using a lower radiation dose, the CT acquisition protocol did not significantly affect attenuation correction and anatomic delineation in PET.
CONCLUSIONS: This study revealed the relation between image information and dose. The current PET/CT imaging acquisition protocol was improved by decreasing the radiation risks without sacrificing the diagnostic values.
METHODS: In this study, the radiation doses delivered from typical germanium-based and CT-based transmission scans were measured and compared using an anthropomorphic Rando Alderson phantom with insertions of thermoluminescent dosimeters. Image geometric distortion and quantified uptake values in PET images with different manipulating CT acquisition protocols for attenuation correction were also evaluated.
RESULTS: It was found that radiation doses during germanium-based transmission scans were almost negligible, while doses from CT-based transmission scans were significantly higher. Using a lower radiation dose, the CT acquisition protocol did not significantly affect attenuation correction and anatomic delineation in PET.
CONCLUSIONS: This study revealed the relation between image information and dose. The current PET/CT imaging acquisition protocol was improved by decreasing the radiation risks without sacrificing the diagnostic values.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app