CLINICAL TRIAL
JOURNAL ARTICLE
MULTICENTER STUDY
Add like
Add dislike
Add to saved papers

Use of the less invasive stabilization system (LISS) in patients with distal femoral (AO33) fractures: a prospective multicenter study.

INTRODUCTION: The Less Invasive Stabilization System (LISS) is an internal fixator that can be inserted percutaneously by means of a minimally invasive surgical approach. This paper presents the use of this system exclusively in patients with AO33 distal femoral fractures, a group in which fracture management is often complicated by multiple trauma and soft-tissue damage.

MATERIALS AND METHODS: A series of 62 patients (mean age 52 years) with 66 AO33A or C fractures received the LISS implant and were followed prospectively for 12 months. Assessments included the radiographic evaluation of implant placement and healing, measurement of valgus/varus and sagittal joint malalignment, and the assessment of weight-bearing capacity and range of motion.

RESULTS: The mean duration of surgery was 85 min (range 40-135 min) for 33A fractures and 149 min (range 50-300 min) for 33C fractures. Postoperative radiographic assessments showed that the LISS implant was positioned correctly in 59 cases (89%). Assessment of valgus/varus alignment showed correct axial alignment in 49 cases (74%), a deviation of 5-10 degrees in 13 cases, and a 10-20 degrees deviation in 1 case. Correct sagittal alignment was observed in 56 cases (85%), 5-10 degrees malalignment in 5 cases, and 10-20 degrees in 2 cases. During the course of the 1-year follow-up, 8 patients (9 fractures) died; 2 other patients were not available for follow-up. Complete fracture healing was achieved in 85% of the followed-up patients. Forty-eight patients were capable of full weight-bearing. After fracture healing, maximum flexion of the injured limb was on average 80% of the range of the uninjured limb. An extension deficit of more than 5 degrees was measured in 3 cases. Of the 62 patients, 14 underwent further operations during the course of follow-up. Of these, 6 required bone grafting and 3 refixation of the implant due to implant loosening. Deep infections requiring several debridements occurred in 2 patients. No complications relating solely to the implant were observed.

CONCLUSION: The LISS showed good overall results in the treatment of these difficult fractures. It is a good alternative to conventional extramedullary and intramedullary stabilizing techniques, especially in more complex fracture situations.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app