Treatment accuracy of fractionated stereotactic radiotherapy.
Radiotherapy and Oncology 2005 January
BACKGROUND AND PURPOSE: To assess the geometric accuracy of the delivery of fractionated stereotactic radiotherapy (FSRT) for brain tumours using the Gill-Thomas-Cosman (GTC) relocatable frame. Accuracy of treatment delivery was measured via portal images acquired with an amorphous silicon based electronic portal imager (EPI). Results were used to assess the existing verification process and to review the current margins used for the expansion of clinical target volume (CTV) to planning target volume (PTV).
PATIENTS AND METHODS: Patients were immobilized in a GTC frame. Target volume definition was performed on localization CT and MRI scans and a CTV to PTV margin of 5mm (based on initial experience) was introduced in 3D. A Brown-Roberts-Wells (BRW) fiducial system was used for stereotactic coordinate definition. The existing verification process consisted of an intercomparison of the coordinates of the isocentres and anatomy between the localization and verification CT scans. Treatment was delivered with 6 MV photons using four fixed non-coplanar conformal fields using a multi-leaf collimator. Portal imaging verification consisted of the acquisition of orthogonal images centred through the treatment isocentre. Digitally reconstructed radiographs (DRRs) created from the CT localization scans were used as reference images. Semi-automated matching software was used to quantify set up deviations (displacements and rotations) between reference and portal images.
RESULTS: One hundred and twenty six anterior and 123 lateral portal images were available for analysis for set up deviations. For displacements, the total errors in the cranial/caudal direction were shown to have the largest SD's of 1.2 mm, while systematic and random errors reached SD's of 1.0 and 0.7 mm, respectively, in the cranial/caudal direction. The corresponding data for rotational errors (the largest deviation was found in the sagittal plane) was 0.7 degrees SD (total error), 0.5 degrees (systematic) and 0.5 degrees (random). The total 3D displacement was 1.8 mm (mean), 0.8 mm (SD) with a range of 0.3-3.9 mm.
CONCLUSIONS: Portal imaging has shown that the existing verification and treatment delivery techniques currently in use result in highly reproducible setups. Random and systematic errors in the treatment planning and delivery chain will always occur, but monitoring and minimising them is an essential component of quality control. Portal imaging provides fast and accurate facility for monitoring patients on treatment and the results of this study have shown that a reduction in CTV to PTV margin from 5 to 4 mm (resulting in a considerable increase in the volume of normal tissue sparing) could be made.
PATIENTS AND METHODS: Patients were immobilized in a GTC frame. Target volume definition was performed on localization CT and MRI scans and a CTV to PTV margin of 5mm (based on initial experience) was introduced in 3D. A Brown-Roberts-Wells (BRW) fiducial system was used for stereotactic coordinate definition. The existing verification process consisted of an intercomparison of the coordinates of the isocentres and anatomy between the localization and verification CT scans. Treatment was delivered with 6 MV photons using four fixed non-coplanar conformal fields using a multi-leaf collimator. Portal imaging verification consisted of the acquisition of orthogonal images centred through the treatment isocentre. Digitally reconstructed radiographs (DRRs) created from the CT localization scans were used as reference images. Semi-automated matching software was used to quantify set up deviations (displacements and rotations) between reference and portal images.
RESULTS: One hundred and twenty six anterior and 123 lateral portal images were available for analysis for set up deviations. For displacements, the total errors in the cranial/caudal direction were shown to have the largest SD's of 1.2 mm, while systematic and random errors reached SD's of 1.0 and 0.7 mm, respectively, in the cranial/caudal direction. The corresponding data for rotational errors (the largest deviation was found in the sagittal plane) was 0.7 degrees SD (total error), 0.5 degrees (systematic) and 0.5 degrees (random). The total 3D displacement was 1.8 mm (mean), 0.8 mm (SD) with a range of 0.3-3.9 mm.
CONCLUSIONS: Portal imaging has shown that the existing verification and treatment delivery techniques currently in use result in highly reproducible setups. Random and systematic errors in the treatment planning and delivery chain will always occur, but monitoring and minimising them is an essential component of quality control. Portal imaging provides fast and accurate facility for monitoring patients on treatment and the results of this study have shown that a reduction in CTV to PTV margin from 5 to 4 mm (resulting in a considerable increase in the volume of normal tissue sparing) could be made.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app