Accuracy of parents in measuring body temperature with a tympanic thermometer.

BMC Family Practice 2005 January 12
BACKGROUND: It is now common for parents to measure tympanic temperatures in children. The objective of this study was to assess the diagnostic accuracy of these measurements.

METHODS: Parents and then nurses measured the temperature of 60 children with a tympanic thermometer designed for home use (home thermometer). The reference standard was a temperature measured by a nurse with a model of tympanic thermometer commonly used in hospitals (hospital thermometer). A difference of >or= 0.5 degrees C was considered clinically significant. A fever was defined as a temperature >or= 38.5 degrees C.

RESULTS: The mean absolute difference between the readings done by the parent and the nurse with the home thermometer was 0.44 +/- 0.61 degrees C, and 33% of the readings differed by >or= 0.5 degrees C. The mean absolute difference between the readings done by the parent with the home thermometer and the nurse with the hospital thermometer was 0.51 +/- 0.63 degrees C, and 72 % of the readings differed by >or= 0.5 degrees C. Using the home thermometer, parents detected fever with a sensitivity of 76% (95% CI 50-93%), a specificity of 95% (95% CI 84-99%), a positive predictive value of 87% (95% CI 60-98%), and a negative predictive value of 91% (95% CI 79-98 %). In comparing the readings the nurse obtained from the two different tympanic thermometers, the mean absolute difference was 0.24 +/- 0.22 degrees C. Nurses detected fever with a sensitivity of 94% (95 % CI 71-100 %), a specificity of 88% (95% CI 75-96 %), a positive predictive value of 76% (95% CI 53-92%), and a negative predictive value of 97% (95%CI 87-100 %) using the home thermometer. The intraclass correlation coefficient for the three sets of readings was 0.80, and the consistency of readings was not affected by the body temperature.

CONCLUSIONS: The readings done by parents with a tympanic thermometer designed for home use differed a clinically significant amount from the reference standard (readings done by nurses with a model of tympanic thermometer commonly used in hospitals) the majority of the time, and parents failed to detect fever about one-quarter of the time. Tympanic readings reported by parents should be interpreted with great caution.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app