COMPARATIVE STUDY
JOURNAL ARTICLE

Biventricular pacing and left ventricular pacing in heart failure: similar hemodynamic improvement despite marked electromechanical differences

Pierre Bordachar, Stephane Lafitte, Sylvain Reuter, Stephane Garrigue, Prashanthan Sanders, Raymond Roudaut, Pierre Jaïs, Michel Haïssaguerre, Jacques Clementy
Journal of Cardiovascular Electrophysiology 2004, 15 (12): 1342-7
15610275

INTRODUCTION: We conducted an acute echocardiographic study comparing hemodynamic and ventricular dyssynchrony parameters during left ventricular pacing (LVP) and biventricular pacing (BVP). We sought to clarify the mechanisms responsible for similar hemodynamic improvement despite differences in electrical activation.

METHODS AND RESULTS: Thirty-three patients underwent echocardiography prior to implantation with a multisite pacing device (spontaneous rhythm [SR]) and 2 days after implantation (BVP and LVP). Interventricular dyssynchrony (pulsed-wave Doppler), extent of myocardium displaying delayed longitudinal contraction (%DLC; tissue tracking), and index of LV dyssynchrony (pulsed-wave tissue Doppler imaging) were assessed. Compared to SR, BVP and LVP caused similar significant improvement of cardiac output (LVP: 3.2 +/- 0.5, BVP: 3.1 +/- 0.7, SR: 2.3 +/- 0.6 L/min; P < 0.01) and mitral regurgitation (LVP: 25.1 +/- 10, BVP: 24.7 +/- 11, baseline: 37.9 +/- 14% jet area/left atria area; P < 0.01). LVP resulted in a smaller index of LV dyssynchrony than BVP (29 +/- 10 vs 34 +/- 14; P < 0.05). However, LVP exhibited a longer aortic preejection delay (220 +/- 34 vs 186 +/- 28 msec; P < 0.01), longer LV electromechanical delays (244.5 +/- 39 vs 209.5 +/- 47 msec; P < 0.05), greater interventricular dyssynchrony (56.6 +/- 18 vs 31.4 +/- 18; P < 0.01), and higher%DLC (40.1 +/- 08 vs 30.3 +/- 09; P < 0.05), leading to shorter LV filling time (387 +/- 54 vs 348 +/- 44 msec; P < 0.05) compared to BVP.

CONCLUSION: Although LVP and BVP provide similar hemodynamic improvement, LVP results in more homogeneous but substantially delayed LV contraction, leading to shortened filling time and less reduction in postsystolic contraction. These data may influence the choice of individual optimal pacing configuration.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
15610275
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"