Screening for diabetic retinopathy using a digital non-mydriatic camera compared with standard 35-mm stereo colour transparencies.
Acta Ophthalmologica Scandinavica 2004 December
PURPOSE: To investigate the use of a digital non-mydriatic camera for determining the ETDRS clinical level of diabetic retinopathy, and to evaluate its use in a screening setting to appropriately determine the need for referral to an ophthalmologist (ETDRS level > or = 35).
METHODS: A total of 83 patients with diabetes were photographed with and without pharmacological pupil dilation at an ophthalmology department using a digital non-mydriatic camera, obtaining two sets of five non-stereoscopic, 45 degree field images of each eye. ETDRS seven standard field, 35-mm stereoscopic colour fundus photographs were also obtained. A subgroup of 59 patients was photographed at an optician's shop using the digital non-mydriatic camera without pupil dilation.
RESULTS: There was substantial agreement between the clinical level of diabetic retinopathy assessed from the 35-mm photographs and the digital images: the ophthalmology department (kappa = 0.76) with pupil dilation and (kappa = 0.66) without pupil dilation, respectively, and at the optician's (kappa = 0.60 without pupil dilation). With respect to the need for referral to an ophthalmologist, there was almost perfect agreement in the ophthalmology department (kappa = 0.88) with pupil dilation and (kappa = 0.84) without pupil dilation, respectively, and those taken at the optician's (kappa = 0.87 without pupil dilation).
CONCLUSION: A digital non-mydriatic camera may be used in a screening situation to appropriately determine the need for referral to an ophthalmologist (ETDRS level > or = 35).
METHODS: A total of 83 patients with diabetes were photographed with and without pharmacological pupil dilation at an ophthalmology department using a digital non-mydriatic camera, obtaining two sets of five non-stereoscopic, 45 degree field images of each eye. ETDRS seven standard field, 35-mm stereoscopic colour fundus photographs were also obtained. A subgroup of 59 patients was photographed at an optician's shop using the digital non-mydriatic camera without pupil dilation.
RESULTS: There was substantial agreement between the clinical level of diabetic retinopathy assessed from the 35-mm photographs and the digital images: the ophthalmology department (kappa = 0.76) with pupil dilation and (kappa = 0.66) without pupil dilation, respectively, and at the optician's (kappa = 0.60 without pupil dilation). With respect to the need for referral to an ophthalmologist, there was almost perfect agreement in the ophthalmology department (kappa = 0.88) with pupil dilation and (kappa = 0.84) without pupil dilation, respectively, and those taken at the optician's (kappa = 0.87 without pupil dilation).
CONCLUSION: A digital non-mydriatic camera may be used in a screening situation to appropriately determine the need for referral to an ophthalmologist (ETDRS level > or = 35).
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app