Measurement of cardiac output before and after cardiopulmonary bypass: Comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing.

OBJECTIVES: A noninvasive continuous cardiac output system (NICO) has been developed recently. NICO uses a ratio of the change in the end-tidal carbon dioxide partial pressure and carbon dioxide elimination in response to a brief period of partial rebreathing to measure CO. The aim of this study was to compare the agreement among NICO, bolus (TDCO), and continuous thermodilution (CCO), with transit-time flowmetry of the ascending aorta using an ultrasonic flow probe (UFP) before and after cardiopulmonary bypass (CPB).

DESIGN: Prospective, observational human study.

SETTING: Veterans Affairs Medical Center Hospital.

PARTICIPANTS: Sixty-eight patients.

METHODS: Matched sets of CO measurements between NICO, TDCO, CCO, and UFP were collected in 68 patients undergoing elective CABG at specific time periods before and after separation from CPB. After anesthetic induction, all patients had an NICO sensor attached between the endotracheal tube and the breathing circuit, a PAC floated into the pulmonary artery for TDCO and CCO monitoring, and a UFP positioned on the ascending aorta and used for the reference CO. Bland-Altman analysis was used to compare the agreement among the different methods.

MEASUREMENTS AND MAIN RESULTS: Bland-Altman analysis of CO measurements before CPB yielded a bias, precision, and percent error of 0.04 L/min +/- 1.07 L/min (44.8%) for NICO, 0.18 L/min +/- 1.01 L/min (41.7%) for TDCO, and 0.29 L/min +/- 1.40 L/min (57.5%) for CCO compared with simultaneous UFP CO measurements, respectively. After separation from CPB (average 29 mins), bias, precision, and percent error were -0.46 L/min +/- 1.06 L/min (37.3%) for NICO, 0.35 L/min +/- 1.39 L/min (46.1%) for TDCO, and 0.36 L/min +/- 1.96 L/min (64.7%) for CCO compared with UFP CO measurements, respectively.

CONCLUSIONS: Before initiation of CPB, the accuracy for all 3 techniques was similar. After separation from CPB, the tendency was for NICO to underestimate CO and for TDCO and CCO to overestimate it. NICO offers an alternative to invasive CO measurement.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app