COMPARATIVE STUDY
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

Clinical experience with a new detection algorithm for differentiation of supraventricular from ventricular tachycardia in a dual-chamber defibrillator.

INTRODUCTION: Inadequate therapy for supraventricular tachyarrhythmias (SVT) is a frequent problem of implantable cardioverter defibrillators (ICD). Dual-chamber ICDs have been developed to improve discrimination of SVT from ventricular tachycardia (VT). We investigated the positive predictivity, sensitivity, and specificity of a new algorithm, the SMART detection trade mark algorithm, incorporated in the Phylax AV (Biotronik) dual-chamber ICD.

METHODS AND RESULTS: Two hundred nine patients (185 men, age 64 +/- 11 years) received a Phylax AV ICD with SMART detection trade mark activated. In 138 of these patients, 1,245 sustained tachycardia episodes with a detailed electrogram were stored in the device during a follow-up period of 10 +/- 6 months. Episodes were correctly classified as ventricular fibrillation (VF, n = 178) in 52 patients, VT (n = 641) in 98 patients, and SVT (n = 385) in 48 patients by the algorithm. Forty-one true SVT episodes (3.3%) were misclassified as VT: atrial fibrillation (n = 7) and flutter (n = 1), sinus tachycardia (n = 12), and other SVT (n = 21). The positive predictivity for VF/VT was 94.5% (95% CI 92.7-95.8) uncorrected and 94.5% (95% CI 92.9-95.8%) corrected with the generalized equation estimation (GEE) method. The positive predictivity for SVT was 100%. The specificity was 88.9% (95% CI 85.6-91.6%) uncorrected and 89.0% (95% CI 85.6-91.6%) corrected with the GEE method with a sensitivity of 100%.

CONCLUSION: The SMART detection trade mark algorithm was safe and reliable for the detection of all ventricular tachycardias. Although its specificity was high, it should be improved with regard to SVT to avoid inappropriate ICD therapies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app