Evaluation of a new transcutaneous bilirubinometer

M Jeffrey Maisels, Enrique M Ostrea, Suzanne Touch, Sarah E Clune, Eugene Cepeda, Elizabeth Kring, Karin Gracey, Cheryl Jackson, Deborah Talbot, Raywin Huang
Pediatrics 2004, 113 (6): 1628-35
15173483

OBJECTIVE: The objective of this study was to evaluate the Minolta/Hill-Rom Air-Shields Transcutaneous Jaundice Meter model JM-103.

METHODS: We studied a convenience sample of 849 newborns > or =35 weeks of gestation in 3 hospitals. These infants had total serum bilirubin (TSB) levels measured on clinical indication, and transcutaneous bilirubin (TcB) levels were obtained within 1 hour of the TSB levels. The population was 59.2% white, 29.8% black, 4.5% East Asian, 3.8% Middle Eastern, 1.6% Indian/Pakistani, and 1.1% Hispanic.

RESULTS: There was a close correlation between TSB and TcB values in all of the population groups: white (n = 503, r =.949); black (n = 253, r =.822); and East Asian, Indian/Pakistani, and Hispanic (n = 93, r =.926). In the black population, the correlation was less close than in the other groups, and differences between the TcB and TSB measurements tended to increase with rising TSB values. JM-103 values differed from TSB values by 3 mg/dL or more in 2% of white, 3.2% of other, and 17.4% of black infants. In these black infants, the JM-103 value was always greater than the TSB value.

CONCLUSIONS: We conclude that TcB measurements using the JM-103 jaundice meter correlate very closely with TSB levels over the range of TSB encountered in this study. Because only 3.3% of our infants had TSB values >15 mg/dL (257 micro mol/L), more data are needed in this range of TSB concentration. The correlation in black infants is not as close as in other groups, but because the tendency in blacks is for the JM-103 to overestimate serum bilirubin levels, dangerous clinical errors are unlikely to occur. The measurement technique is rapid and simple, and it is easy to perform repeated measurements over time, thus reducing the likelihood of error. TcB measurements with the JM-103 jaundice meter should obviate the need for most serum bilirubin levels in newborn infants > or =35 weeks of gestation, although serum bilirubin measurements are still required when treatment with phototherapy or exchange transfusion is being considered.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
15173483
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"