We have located links that may give you full text access.
COMPARATIVE STUDY
EVALUATION STUDIES
JOURNAL ARTICLE
Left ventricular assist device malfunction: a systematic approach to diagnosis.
OBJECTIVES: A protocol was designed to diagnose the common malfunctions of a left ventricular assist device (LVAD).
BACKGROUND: Mechanical circulatory support, primarily with an LVAD, is increasingly used for treatment of advanced heart failure (HF). Left ventricular assist device dysfunction is a recognized complication; but heretofore, a systematic method to accurately diagnose LVAD dysfunction has not been thoroughly described.
METHODS: We developed a catheter-based protocol designed to characterize a normally functioning LVAD and diagnose multiple types of dysfunction. A total of 15 studies of 10 patients supported with an LVAD were reviewed. All patients had been evaluated due to concerns regarding LVAD dysfunction.
RESULTS: Of 15 examinations performed, 11 documented severe LVAD inflow valve regurgitation. One of these cases proved to have coexistent severe mitral valve regurgitation. One case was diagnosed with distortion of the LVAD outflow graft. One case of suspected embolization from the pumping chamber excluded the outflow graft as the source of emboli. One study had aortic insufficiency.
CONCLUSIONS: As LVAD use for treatment of end-stage HF becomes widespread and durations of support are extended, dysfunction will be increasingly prevalent. This catheter-based protocol provided a practical method to diagnose multiple causes of LVAD dysfunction.
BACKGROUND: Mechanical circulatory support, primarily with an LVAD, is increasingly used for treatment of advanced heart failure (HF). Left ventricular assist device dysfunction is a recognized complication; but heretofore, a systematic method to accurately diagnose LVAD dysfunction has not been thoroughly described.
METHODS: We developed a catheter-based protocol designed to characterize a normally functioning LVAD and diagnose multiple types of dysfunction. A total of 15 studies of 10 patients supported with an LVAD were reviewed. All patients had been evaluated due to concerns regarding LVAD dysfunction.
RESULTS: Of 15 examinations performed, 11 documented severe LVAD inflow valve regurgitation. One of these cases proved to have coexistent severe mitral valve regurgitation. One case was diagnosed with distortion of the LVAD outflow graft. One case of suspected embolization from the pumping chamber excluded the outflow graft as the source of emboli. One study had aortic insufficiency.
CONCLUSIONS: As LVAD use for treatment of end-stage HF becomes widespread and durations of support are extended, dysfunction will be increasingly prevalent. This catheter-based protocol provided a practical method to diagnose multiple causes of LVAD dysfunction.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app