COMPARATIVE STUDY
JOURNAL ARTICLE

Comparison of esophageal Doppler, pulse contour analysis, and real-time pulmonary artery thermodilution for the continuous measurement of cardiac output

Berthold Bein, Frank Worthmann, Peter H Tonner, Andrea Paris, Markus Steinfath, Jürgen Hedderich, Jens Scholz
Journal of Cardiothoracic and Vascular Anesthesia 2004, 18 (2): 185-9
15073709

OBJECTIVE: Continuous measurement of cardiac output (CCO) is of great importance in the critically ill. However, pulmonary artery thermodilution has been questioned for possible complications associated with right heart catheterization. Furthermore, measurements are delayed in the continuous mode during rapid hemodynamic changes. A new pulmonary artery catheter CCO device (Aortech, Bellshill, Scotland) enabling real-time update of cardiac output was compared with 2 different, less-invasive methods of CCO determination, esophageal Doppler and pulse contour analysis.

DESIGN: Prospective, observational study.

SETTING: University hospital, single institution.

PARTICIPANTS: Patients scheduled for elective coronary artery bypass grafting (CABG).

INTERVENTIONS: None.

MEASUREMENTS AND MAIN RESULTS: CCO measurements were analyzed using a Bland-Altman plot. Bias between CCO and pulse contour cardiac output (PCCO), and Doppler-derived cardiac output (UCCO) was (mean +/- 1 SD) -0.71 +/- 1 L/min versus -0.15 +/- 1.09 L/min, and between UCCO and PCCO -0.58 +/- 1.06 L/min. Bias was not significantly different among methods, nor were comparative values before and after cardiopulmonary bypass (p > 0.05).

CONCLUSIONS: Agreement between the CCO method and both less-invasive measurements was clinically acceptable. There were no adverse events associated with the use of either device.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
15073709
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"