COMPARATIVE STUDY
EVALUATION STUDIES
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

In vitro performance characteristics of valved holding chamber and spacer devices with a fluticasone metered-dose inhaler.

Pharmacotherapy 2004 Februrary
STUDY OBJECTIVE: To compare the in vitro aerosol deposition characteristics of several commercially available valved holding chamber (VHC) and spacer devices used with a fluticasone metered-dose inhaler (MDI).

DESIGN: In vitro aerosol deposition study

SETTING: University-affiliated research center.

DEVICES: Seven VHC devices: BreatheRite, E-Z Spacer, EasiVent, AeroChamber, InspirEase, OptiChamber, and Space Chamber. Six spacer devices: OptiHaler, Aerosol Cloud Enhancer (ACE), Gentle-Haler, MediSpacer, Ellipse, and a 6-inch tube (1-inch inside diameter).

INTERVENTION: The respirable dose (aerosol particles 1-5 microm) of fluticasone was determined by sampling 10 220-microg actuations from five runs with each spacer or VHC plus MDI combination, by using a well-established in vitro cascade impactor method.

MEASUREMENTS AND MAIN RESULTS: Fluticasone aerosol was washed from the impactor with methanol and quantified by means of high-performance liquid chromatography. Differences among outcomes were determined with analysis-of-variance testing. Among spacers, Ellipse had the highest respirable dose (104 microg, p < 0.01). Respirable doses for the 6-inch tube (74.3 microg), Gentle-Haler (81.7 microg), and MediSpacer (82.6 microg) were no different from that of the MDI (p > 0.05), whereas respirable doses of OptiHaler (44.6 microg) and ACE (47.2 microg) were less than those of all other spacers (p < 0.001). Among VHC devices, respirable doses from EasiVent (35.6 microg), AeroChamber (47.0 microg), InspirEase (52.7 microg), OptiChamber (53.1 microg), and Space Chamber (58.3 microg) were not different (p > 0.05), whereas BreatheRite (13.1 microg) and E-Z Spacer (27.3 microg) respirable doses were less than those of the other VHC devices (p < 0.05).

CONCLUSION: Spacers and VHC devices available in the United States do not demonstrate equivalent in vitro performance with the fluticasone MDI. The difference between highest and lowest respirable doses in each device category would likely lead to clinically relevant differences in the quantity of fluticasone delivered to a patient.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app