Analysis of contact mechanics in McKee-farrar metal-on-metal hip implants.

Contact mechanics analysis for a typical McKee-Farrar metal-on-metal hip implant was carried out in this study. The finite element method was used to predict the contact area and the contact pressure distribution at the bearing surfaces. The study investigated the effects of the cement and underlying bone, the geometrical parameters such as the radial clearance between the acetabular cup and the femoral head, and the acetabular cup thickness, as well as other geometrical features on the acetabular cup such as lip and studs. For all the cases considered, the predicted contact pressure distribution was found to be significantly different from that based upon the classical Hertz contact theory, with the maximum value being away from the centre of the contact region. The lip on the cup was found to have a negligible effect on the predicted contact pressure distribution. The presence of the studs on the outside of the cup caused a significant increase in the local contact pressure distribution, and a slight decrease in the contact region. Reasonably good agreement of the predicted contact pressure distribution was found between a three-dimensional anatomical model and a simple two-dimensional axisymmetric model. The interfacial boundary condition between the acetabular cup and the underlying cement, modelled as perfectly fixed or perfectly unbonded, had a negligible effect on the predicted contact parameters. For a given radial clearance of 0.079 mm, the decrease in the thickness of the acetabular cup from 4.5 to 1.5 mm resulted in an increase in the contact half angle from 15 degrees to 26 degrees, and a decrease in the maximum contact pressure from 55 to 20 MPa. For a given acetabular cup thickness of 1.5 mm, a decrease in the radial clearance from 0.158 to 0.0395 mm led to an increase in the contact half-angle from 20 degrees to 30 degrees, and a decrease in the maximum contact pressure from 30 to 10 MPa. For zero clearance, although the contact pressure was significantly reduced over most of the contact area, the whole acetabular cup came into contact with the femoral head, leading to stress concentration at the edge of the cup. Design optimization of the geometrical parameters, in terms of the acetabular cup thickness and the radial clearance, is important, not only to minimize the contact stress at the bearing surfaces, but also to avoid equatorial and edge contact.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app