Radiation doses in interventional radiology procedures: the RAD-IR study: part I: overall measures of dose

Donald L Miller, Stephen Balter, Patricia E Cole, Hollington T Lu, Beth A Schueler, Michael Geisinger, Alejandro Berenstein, Robin Albert, Jeffrey D Georgia, Patrick T Noonan, John F Cardella, James St George, Eric J Russell, Tim W Malisch, Robert L Vogelzang, George L Miller, Jon Anderson
Journal of Vascular and Interventional Radiology: JVIR 2003, 14 (6): 711-27
12817038

PURPOSE: To determine patient radiation doses for interventional radiology and neuroradiology procedures, to identify procedures associated with higher radiation doses, and to determine the effects of various parameters on patient doses.

MATERIALS AND METHODS: A prospective observational study was performed at seven academic medical centers. Each site contributed demographic and radiation dose data for subjects undergoing specific procedures in fluoroscopic suites equipped with built-in cumulative dose (CD) and dose-area-product (DAP) measurement capability compliant with International Electrotechnical Commission standard 60601-2-43. The accuracy of the dosimetry was confirmed by comprehensive measurements and by frequent consistency checks performed over the course of the study.

RESULTS: Data were collected on 2,142 instances of interventional radiology procedures, 48 comprehensive physics evaluations, and 581 periodic consistency checks from the 12 fluoroscopic units in the study. There were wide variations in dose and statistically significant differences in fluoroscopy time, number of images, DAP, and CD for different instances of the same procedure, depending on the nature of the lesion, its anatomic location, and the complexity of the procedure. For the 2,142 instances, observed CD and DAP correlate well overall (r = 0.83, P <.000001), but correlation in individual instances is poor. The same is true for the correlation between fluoroscopy time and CD (r = 0.79, P <.000001). The correlation between fluoroscopy time and DAP (r = 0.60, P <.000001) is not as good. In 6% of instances (128 of 2,142), which were principally embolization procedures, transjugular intrahepatic portosystemic shunt (TIPS) procedures, and renal/visceral artery stent placements, CD was greater than 5 Gy.

CONCLUSIONS: Most procedures studied can result in clinically significant radiation dose to the patient, even when performed by trained operators with use of dose-reducing technology and modern fluoroscopic equipment. Embolization procedures, TIPS creation, and renal/visceral artery stent placement are associated with a substantial likelihood of clinically significant patient dose. At minimum, patient dose data should be recorded in the medical record for these three types of procedures. These data should include indicators of the risk of deterministic effects as well as the risk of stochastic effects.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Remove bar
Read by QxMD icon Read
12817038
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"

We want to hear from doctors like you!

Take a second to answer a survey question.