Evaluation Studies
Journal Article
Add like
Add dislike
Add to saved papers

Feasibility of insertion/implantation of 2.0-mm-diameter gold internal fiducial markers for precise setup and real-time tumor tracking in radiotherapy.

PURPOSE: To examine the feasibility and reliability of insertion of internal fiducial markers into various organs for precise setup and real-time tumor tracking in radiotherapy (RT).

MATERIALS AND METHODS: Equipment and techniques for the insertion of 2.0-mm-diameter gold markers into or near the tumor were developed for spinal/paraspinal lesions, prostate tumors, and liver and lung tumors. Three markers were used to adjust the center of the mass of the target volume to the planned position in spinal/paraspinal lesions and prostate tumors (the three-marker method). The feasibility of the marker insertion and the stability of the position of markers were tested using stopping rules in the clinical protocol (i.e., the procedure was abandoned if 2 of 3 or 3 of 6 patients experienced marker dropping or migration). After the evaluation of the feasibility, the stability of the marker positions was monitored in those patients who entered the dose-escalation study.

RESULTS: Each of the following was shown to be feasible: bronchoscopic insertion for the peripheral lung; image-guided transcutaneous insertion for the liver; cystoscopic and image-guided percutaneous insertion for the prostate; and surgical implantation for spinal/paraspinal lesions. Transcutaneous insertion of markers for spinal/paraspinal lesions and bronchoscopic insertion for central lung lesions were abandoned. Overall, marker implantation was successful and was used for real-time tumor tracking in RT in 90 (90%) of 100 lesions. No serious complications related to the marker insertion were noted for any of the 100 lesions. Using three markers surgically implanted into the vertebral bone, the mean +/- standard deviation in distance among the three markers was within 0.2 +/- 0.6 mm (range -1.4 to 0.8) through the treatment period of 30 days. The distance between the three markers gradually decreased during RT in five of six prostate cancers, consistent with a mean rate of volume regression of 9.3% (range 0.015-13%) in 10 days.

CONCLUSIONS: Internal 2.0-mm-diameter gold markers can be safely inserted into various organs for real-time tumor tracking in RT using the prescribed equipment and techniques. The three-marker method has been shown to be a useful technique for precise setup for spinal/paraspinal lesions and prostate tumors.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app