COMPARATIVE STUDY
JOURNAL ARTICLE
Resurfacing knee arthroplasty in patients with allergic sensitivity to metals.
BACKGROUND: The problem of metal sensitivity (Ni, Cr, and Co) in arthroplasty is still unsolved. To prevent the risk of allergy in cases with proved metal allergy in an epicutaneous test, a Natural Knee total knee arthroplasty, made totally from titanium (Ti-6Al-4V) and polyethylene, can be implanted. The results of this device have to be compared to the results of normal knee arthroplasty designed with a femoral component out of CrCoNi alloy. The mechanical resistance of this alloy is much higher than that of titanium, and therefore it is the standard for the femoral components of most knee-resurfacing devices.
METHODS: Thirty-five patients with a titanium Natural Knee were examined retrospectively after a mean follow-up of 2 years and 5 months (range 6 months to 5 years and 4 months). In comparison, 36 patients with a Genesis-I knee with a CrCoNi alloy femoral component were studied after a mean follow-up of 2 years and 4 months (range 8 months to 6 years and 2 months).
RESULTS: In spite of the lower mechanical resistance of titanium, the Natural Knee showed better results (knee score 84.1 points, function score 77.7 points, HSS score 80.1 points, 82.9% excellent and good results) than the Genesis-I knee (knee score 80.6 points, function score 76.4 points, HSS score 76.4 points, 68.5% excellent and good results), although these differences had no statistical significance.
CONCLUSIONS: The titanium Natural Knee prosthesis has proven to be a reliable knee joint replacement in the medium term.
METHODS: Thirty-five patients with a titanium Natural Knee were examined retrospectively after a mean follow-up of 2 years and 5 months (range 6 months to 5 years and 4 months). In comparison, 36 patients with a Genesis-I knee with a CrCoNi alloy femoral component were studied after a mean follow-up of 2 years and 4 months (range 8 months to 6 years and 2 months).
RESULTS: In spite of the lower mechanical resistance of titanium, the Natural Knee showed better results (knee score 84.1 points, function score 77.7 points, HSS score 80.1 points, 82.9% excellent and good results) than the Genesis-I knee (knee score 80.6 points, function score 76.4 points, HSS score 76.4 points, 68.5% excellent and good results), although these differences had no statistical significance.
CONCLUSIONS: The titanium Natural Knee prosthesis has proven to be a reliable knee joint replacement in the medium term.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app