COMPARATIVE STUDY
JOURNAL ARTICLE
RANDOMIZED CONTROLLED TRIAL

Long-term mechanical ventilation with hygroscopic heat and moisture exchangers used for 48 hours: a prospective clinical, hygrometric, and bacteriologic study

Alexandre Boyer, Guillaume Thiéry, Serge Lasry, Etienne Pigné, Amar Salah, Arnaud de Lassence, Didier Dreyfuss, Jean-Damien Ricard
Critical Care Medicine 2003, 31 (3): 823-9
12626991

OBJECTIVE: To determine whether use of a hygroscopic heat and moisture exchanger (HME) for 48 hrs without change affects its efficiency and the level of bacterial colonization in long-term mechanically ventilated medical intensive care unit patients.

DESIGN: Prospective, randomized clinical study evaluating two hygroscopic HMEs.

SETTING: Medical intensive care unit at a university teaching hospital.

PATIENTS: Long-term mechanically ventilated medical intensive care unit patients, including chronic obstructive pulmonary disease patients.

INTERVENTIONS: Patients were randomly allocated to one of the two HMEs studied (Hygrolife and EdithFlex) and changed every 48 hrs. Devices in both groups could be changed if hygrometric measurements indicated insufficient humidity delivery.

MEASUREMENTS AND MAIN RESULTS: Daily measurements of inspired gas temperature and relative and absolute humidity. In addition, cultures of tracheal aspirations and both patient and ventilator sides of the device were performed after 48 hrs of use. Ventilatory variables and clinical indicators of efficient humidification were also recorded. Prolonged use of both HMEs was safe and efficient (no tracheal tube occlusion occurred). Mean duration of mechanical ventilation was 20 days. Both clinical indicators and hygrometric measurements showed that both devices performed well during 48 hrs. Absolute humidity with EdithFlex was significantly higher on day 0 and day 1 than with Hygrolife. Absolute humidity measured in chronic obstructive pulmonary disease patients was identical to that measured in the rest of the study population. Tracheal colonization and HME colonization were similar with both HMEs. Bacterial contamination of the ventilator side of both devices was markedly low.

CONCLUSIONS: These two purely hygroscopic HMEs provided safe and efficient humidification during a 48-hr period of use in long-term mechanically ventilated medical intensive care unit patients, including chronic obstructive pulmonary disease patients. In addition, they maintained ventilatory circuits clean, despite the absence of filtering media. The cost of mechanical ventilation is consequently reduced.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Remove bar
Read by QxMD icon Read
12626991
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"