COMPARATIVE STUDY
JOURNAL ARTICLE

[Biodegradable meniscus fixations: a comparative biomechanical study]

R Seil, S Rupp, C Jurecka, T Georg, D Kohn
Revue de Chirurgie Orthopédique et Réparatrice de L'appareil Moteur 2003, 89 (1): 35-43
12610434

PURPOSE: Numerous new all inside meniscus repair techniques using biodegradable fixation devices have become available in the last years. Few studies have dealt with the biomechanical properties of these implants so far. The aim of this study was to analyze meniscus repairs under cyclic loading conditions for different fixation devices (Meniscus Arrow, Clearfix Meniscal Screw, Meniscal Dart, BioStinger) and to compare them to standard suture techniques.

MATERIAL AND METHODS: Eighty meniscus repairs on 27 medial porcine menisci were performed. For biomechanical testing loosening, number of failures, failure modes and failure loads were evaluated under cyclic and maximum loading conditions. Testing was performed on a materials testing machine.

RESULTS: During cyclic loading 10 out of 40 (25%) fixation devices failed. A gap appeared between the 2 parts of the meniscus within the first loading cycles. The failure strengths obtained with the 4 meniscus fixation devices did not differ significantly (Meniscus Arrow: 44 N; Clearfix Meniscal Screw: 35 N; Meniscal Dart: 33 N; BioStinger: 54 N). Besides for the BioStinger, the values of the sutures were significantly higher. The use of a stronger suture material (PDS 0) increased the failure load significantly. During maximum loading the fixation devices failed by both pull-out of the head or the tip of the device, except for the BioStinger showing only pull-out of the head. All sutures failed by breakage, except for the PDS 0 horizontal sutures in which 50% of complete tissue failures could be noted.

CONCLUSION: This study shows that meniscal fixation devices may fail under repetitive loading conditions. A gap appeared between the meniscal margins within the first loading cycles irrespective of the suture type and fixation device used. Both gapping and failure modes of the fixation devices may be design-specific. The maximum failure loads differed significantly between the bioabsorbable devices and the sutures. Stronger suture materials increased the failure loads significantly. To allow a complete evaluation of new meniscus fixation devices not only biomechanical, but also biological and clinical parameters need to be analyzed in forthcoming studies.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Remove bar
Read by QxMD icon Read
12610434
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"