COMPARATIVE STUDY
IN VITRO
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
RESEARCH SUPPORT, U.S. GOV'T, NON-P.H.S.
RESEARCH SUPPORT, U.S. GOV'T, P.H.S.
Add like
Add dislike
Add to saved papers

Reduction in lateral thermal damage using heat-conducting templates: a comparison of continuous wave and pulsed CO2 lasers.

BACKGROUND AND OBJECTIVES: The advantages of the continuous wave (c.w.) CO(2) laser are offset by the delay in laser wound healing secondary to thermal damage. We have developed novel heat-conducting templates to reduce laser thermal damage. Because shortened pulse durations also decrease thermal damage, we tested the effectiveness of heat-conducting templates with a c.w. CO(2) clinical laser and a short-pulsed CO(2) laser to determine the best method and mechanism to minimize thermal damage.

STUDY DESIGN/MATERIALS AND METHODS: Comparison of 0.2-second shuttered c.w. and 5-microsecond pulsed CO(2) lasers were made by doing incisions on 150 tissue samples from reduction mammoplasties and abdominoplasties. Copper, aluminum, glass, and Plexiglass heat-conducting templates were tested against no template (air) with both lasers. Histological samples were evaluated using computerized morphometrics analysis.

RESULTS: Statistically significant reductions in lateral thermal damage were seen with the copper (50%) and aluminum (39%) templates used with the c.w. CO(2) laser. Only the copper template (39%) significantly reduced thermal damage when used with the pulsed CO(2) laser. Less thermal damage was seen using the pulsed CO(2) laser compared to the c.w. CO(2) laser with each template.

CONCLUSIONS: Heat-conducting templates significantly reduced the amount of lateral thermal damage when used with the c.w. CO(2) laser (copper and aluminum) and short-pulsed CO(2) laser (copper). The c.w. CO(2) laser with the copper template compared favorably to the short-pulsed CO(2) laser without a template. Therefore, both heat conductive templates and short-pulse structure provide successful methods for reducing lateral thermal damage, and a combination of the two appears to provide optimal results.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app