Evaluation of 4-mm to 7-mm versus 6-mm prosthetic brachial-antecubital forearm loop access for hemodialysis: results of a randomized multicenter clinical trial.

OBJECTIVE: Prosthetic arteriovenous fistulas for hemodialysis vascular access have a high incidence rate of thrombotic occlusions that result in graft failure. This randomized multicenter study was performed to assess the patency rates and the effect of 4-mm to 7-mm grafts on the development of stenoses.

METHODS: A total of 109 patients who needed vascular access for hemodialysis were randomized to receive either 6-mm (n = 57) or 4-mm to 7-mm prosthetic brachial-antecubital forearm loop accesses (polytetrafluoroethylene). Duplex scanning, with measurement of blood flow and peak systolic velocity and detection of stenoses (>50%), was performed at 1, 6, and 12 months after surgery. Clinical data were obtained in a prospective manner and primary, assisted primary, and secondary patency rates were calculated with the Kaplan-Meier life-table analysis. Statistical analysis was performed with the independent samples t test and chi(2) test.

RESULTS: At 1 year, patency rates were similar for both 4-mm to 7-mm and 6-mm prostheses (primary, 46% versus 43%; assisted primary, 62% versus 58%; secondary, 87% versus 91%). The incidence rate of thrombotic occlusion was comparable for both groups (0.74/patient-year versus 0.88/patient-year; P >.05). Mean graft flow at 1, 6, and 12 months was 1416 versus 1415 mL/min, 1345 versus 1319 mL/min, and 1595 versus 1265 mL/min (P >.05) for 4-mm to 7-mm and 6-mm grafts, respectively. Also, no differences in peak systolic velocities in any part of the grafts were observed. The percentage of stenoses detected was equal in both groups at 1 year after surgery (27% versus 20%; P >.05).

CONCLUSION: A 4-mm to 7-mm tapered prosthetic brachial-antecubital forearm loop access did not reduce the incidence rates of stenoses and thrombotic occlusions compared with a 6-mm prosthetic conduit. Moreover, no differences in patency rates were observed. Therefore, we believe that the 4-mm to 7-mm graft should not be used routinely for hemodialysis vascular access.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app