Discrepant sensitivity of thromboplastin reagents to clotting factor levels explored by the prothrombin time in patients on stable oral anticoagulant treatment: impact on the international normalized ratio system.
Haematologica 2002 December
BACKGROUND AND OBJECTIVES: We tested the principle of local International Normalized Ratio (INR) calibration using INR calibrator plasmas (PT Calibration Plasma Kit, Behring), two thomboplastin reagents (Neoplastin plus, rabbit brain, Stago, and Recombiplastin, recombinant human tissue factor, Ortho Diagnostics) and the same coagulometer (STA, Stago) on 92 patients on stable oral anticoagulant treatment.
DESIGN AND METHODS: A four-point calibration was obtained with each reagent by linear regression (sec/INR) on a log-log scale (r > or = 0.999). The bias between the two reagents (Recombiplastin - Neoplastin Plus) was reduced from 31.7% to 17.5% and 7.5% (p=0.001) when results were expressed, respectively, as PT ratio (using the mean normal prothrombin time as denominator term), INR (using instrument-specific ISI supplied by the manufacturers) and calibrated INR, but there was a consistently significant regression of the differences over the average values even after log transformation (r > or = 0.586). The bias between the reagents was reduced to 1% (p=ns) when assuming Recombiplastin as the reference thromboplastin and applying Tomenson's correction, but limits of agreements were as large as 20%. Factor VII, X, V and II activity was measured with the two thromboplastin reagents in all plasma samples using immunodepleted plasmas (Stago).
RESULTS: Statistically significant biases were observed for all clotting factors with the two reagents (Recombiplastin Neoplastin Plus) and ranged from 3.5 % (FII) to 37.2% (FVII). In addition, for FVII and FV there was a significant regression of the difference over the average value (after log-transformation, r > or = 0.282). The patients were divided into 3 groups according to their degree of anticoagulation (INR <2.0; INR between 2.0 and 3.5; INR >3.5). Factor levels differed significantly with the two reagents throughout the 3 groups of patients. In addition, the relative distributions of the 3 vitamin K-dependent factors also differed in the 3 groups with the two thromboplastin reagents.
INTERPRETATION AND CONCLUSIONS: The discrepant sensitivity to factor VII, X and V levels of the two thromboplastin reagents explored in this study prevents INR calibration with commercially available calibrator plasmas and is responsible for a significant variability in INR values even under optimal conditions of INR calibration.
DESIGN AND METHODS: A four-point calibration was obtained with each reagent by linear regression (sec/INR) on a log-log scale (r > or = 0.999). The bias between the two reagents (Recombiplastin - Neoplastin Plus) was reduced from 31.7% to 17.5% and 7.5% (p=0.001) when results were expressed, respectively, as PT ratio (using the mean normal prothrombin time as denominator term), INR (using instrument-specific ISI supplied by the manufacturers) and calibrated INR, but there was a consistently significant regression of the differences over the average values even after log transformation (r > or = 0.586). The bias between the reagents was reduced to 1% (p=ns) when assuming Recombiplastin as the reference thromboplastin and applying Tomenson's correction, but limits of agreements were as large as 20%. Factor VII, X, V and II activity was measured with the two thromboplastin reagents in all plasma samples using immunodepleted plasmas (Stago).
RESULTS: Statistically significant biases were observed for all clotting factors with the two reagents (Recombiplastin Neoplastin Plus) and ranged from 3.5 % (FII) to 37.2% (FVII). In addition, for FVII and FV there was a significant regression of the difference over the average value (after log-transformation, r > or = 0.282). The patients were divided into 3 groups according to their degree of anticoagulation (INR <2.0; INR between 2.0 and 3.5; INR >3.5). Factor levels differed significantly with the two reagents throughout the 3 groups of patients. In addition, the relative distributions of the 3 vitamin K-dependent factors also differed in the 3 groups with the two thromboplastin reagents.
INTERPRETATION AND CONCLUSIONS: The discrepant sensitivity to factor VII, X and V levels of the two thromboplastin reagents explored in this study prevents INR calibration with commercially available calibrator plasmas and is responsible for a significant variability in INR values even under optimal conditions of INR calibration.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app