Cardiac output measurement during cardiac surgery: esophageal Doppler versus pulmonary artery catheter.

BACKGROUND: Bolus thermodilution cardiac output (BCO) measurement has been considered as the "gold standard" for cardiac output (CO) measurement. However, it requires placement of a pulmonary artery (PA) catheter, and questions have been raised regarding the risk/benefit ratio of this invasive technique. Furthermore, great variations between measurements have been reported. Continuous thermodilution CO (CCO) measurement is reported to be a better alternative, but it still requires the placement of a PA catheter. Esophageal echo-Doppler ultrasonography (ED) provides non-invasive continuous measurement of CO (ED-CO). This study was thus designed to compare the agreement between ED-CO and both thermodilution techniques (BCO and CCO).

METHODS: Twenty-four patients undergoing primary coronary artery bypass graft surgery were randomized to have a PA catheter placed for measurement of either BCO or CCO. All patients also had an ED probe placed. In Group I patients (n = 12), BCO measurement was carried out every 15 minutes throughout the surgery except during cardiopulmonary bypass, with concurrent ED-CO reading recorded at the same time point. In Group II patients (n = 12), CCO and ED-CO measurements were recorded at the same designated points of time as in Group I. The agreement between methods (BCO vs. ED-CO or CCO vs. ED-CO) was assessed using Bland-Altman method.

RESULTS: The range of measured CO of each method was 2.1 to 9.4 l/min for BCO, 2.4 to 9.2 l/min for CCO and 2.3 to 8.9 l/min for ED-CO. ED-CO and CCO had excellent agreement with a linear regression coefficient (r2 value) of 0.846, and a bias (mean difference) and SD of bias of 0.05 +/- 0.49 l/min. In contrast, the agreement between BCO and ED-CO was poorer; correlation was low (r2 value 0.406) and both the bias and SD of bias were high (0.11 +/- 1.12 l/min). Furthermore, BCO measurements had poor reproducibility, whereas both ED-CO and CCO measurements had good reproducibility.

CONCLUSIONS: Esophageal echo-Doppler ultrasonography is a satisfactory alternative for cardiac output measurement because it gives a value in good agreement with CCO measurement. With significant between-measurement variations, the accuracy and precision of BCO are uncertain, and it should not be considered as the "gold standard".

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app