[A comparative study of the follow-up and hemodynamics in vivo of 21 mm Carpentier-Edwards supra-annular and perimount bioprostheses].
Revista Española de Cardiología 2002 July
INTRODUCTION AND OBJECTIVES: Analysis and comparison of the clinical performance and hemodynamics in vivo of 21 mm Carpentier-Edwards supra-annular (CESA) and Perimount (CEPM) aortic bioprostheses.
METHODS: A follow-up study was made of 40 patients implanted a 21 mm CESA (n = 21) or CEPM (n = 19) prosthesis between October 1992 and September 1997. All eligible survivors (14 CESA, 12 CEPM) were assessed echocardiographically.
RESULTS: There were no significant differences between models in the effective orifice area (1.6 cm2 for CESA, 1.44 cm2 for CEPM), peak flow rate (rest: 2.5 m/s for CESA, 2.3 m/s for CEPM; post-dobutamine: 3.4 m/s for CESA, 3.3 m/s for CEPM), mean flow rate (rest: 1.7 m/s for CESA, 1.6 m/s for CEPM; post-dobutamine: 2.5 m/s for CESA, 2.2 m/s for CEPM), peak gradient (rest: 28.3 mmHg for CESA, 21.6 mmHg for CEPM; post-dobutamine: 48.4 mmHg for CESA, 41.6 mmHg for CEPM), and mean gradient (rest: 15.8 mmHg for CESA, 12.0 mmHg for CEPM; post-dobutamine: 28.5 mmHg for CESA, 22.5 mmHg for CEPM).
CONCLUSION: In our experience, these two prosthetic models have similar hemodynamic characteristics in small aortic annuli.
METHODS: A follow-up study was made of 40 patients implanted a 21 mm CESA (n = 21) or CEPM (n = 19) prosthesis between October 1992 and September 1997. All eligible survivors (14 CESA, 12 CEPM) were assessed echocardiographically.
RESULTS: There were no significant differences between models in the effective orifice area (1.6 cm2 for CESA, 1.44 cm2 for CEPM), peak flow rate (rest: 2.5 m/s for CESA, 2.3 m/s for CEPM; post-dobutamine: 3.4 m/s for CESA, 3.3 m/s for CEPM), mean flow rate (rest: 1.7 m/s for CESA, 1.6 m/s for CEPM; post-dobutamine: 2.5 m/s for CESA, 2.2 m/s for CEPM), peak gradient (rest: 28.3 mmHg for CESA, 21.6 mmHg for CEPM; post-dobutamine: 48.4 mmHg for CESA, 41.6 mmHg for CEPM), and mean gradient (rest: 15.8 mmHg for CESA, 12.0 mmHg for CEPM; post-dobutamine: 28.5 mmHg for CESA, 22.5 mmHg for CEPM).
CONCLUSION: In our experience, these two prosthetic models have similar hemodynamic characteristics in small aortic annuli.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app