Runaway pulse generator malfunction resulting from undetected battery depletion.

Runaway pacemaker is an uncommon, potentially lethal circuit malfunction characterized by sudden onset of erratic pacing at rapid nonphysiological rates. Two patients with a single chamber pacemaker (Medtronic ST 8331 and 8419) presented with episodic dizziness. ECG revealed recurrent decrescendo amplitude episodes of runaway stimuli at 2,400 and 2,600 ppm, approximately 3 seconds in duration, separated by pacing at 62.5 and 65 ppm, respectively. Fortunately the runaway stimuli were subthreshold and did not result in capture of the ventricle. Emergency pulse generator replacement was uneventful. Both leads were normal and both pulse generators had low battery voltages at 1.488 and 1.78 V, respectively.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app