Add like
Add dislike
Add to saved papers

Transvenous transjugular renal core biopsy with a redesigned biopsy set including a blunt-tipped needle.

A novel 19-gauge, blunt-tipped, side cutting single throw, 70-cm long transjugular needle, specifically designed for transvenous kidney biopsy, was used in seven patients with high risk for bleeding. A mean of 4 device-passes (3-6) per patient resulted in a satisfactory specimen for pathological diagnosis. Immediate post-biopsy nonenhanced CT was performed to evaluate for bleeding at the biopsy site. All patients were observed for 2 hr after the procedure. No clinically significant immediate or late complication was noted.