Immersion A-scan compared with partial coherence interferometry: outcomes analysis.
Journal of Cataract and Refractive Surgery 2002 Februrary
PURPOSE: To compare 2 methods of axial length measurement, immersion ultrasonography and partial coherence interferometry, and to elucidate surgical outcomes based on immersion measurements.
SETTING: Oregon Eye Institute, Eugene, Oregon, USA.
METHODS: Axial length measurements in 50 cataractous eyes were obtained by optical biometry (IOLMaster, Zeiss Humphrey Systems) and immersion ultrasound (Axis II, Quantel Medical), and the results were compared. Intraocular lens (IOL) power calculations in the same eyes after cataract extraction and posterior chamber IOL implantation were evaluated retrospectively based on the postoperative spherical equivalent prediction error.
RESULTS: Immersion ultrasonography and partial coherence interferometry measurements correlated in a highly positive manner (correlation coefficient = 0.996). Outcomes analysis demonstrated 92.0% of eyes were within +/-0.5 diopter of emmetropia based on immersion axial length measurements.
CONCLUSION: Immersion ultrasonography provided highly accurate axial length measurements and permitted highly accurate IOL power calculations.
SETTING: Oregon Eye Institute, Eugene, Oregon, USA.
METHODS: Axial length measurements in 50 cataractous eyes were obtained by optical biometry (IOLMaster, Zeiss Humphrey Systems) and immersion ultrasound (Axis II, Quantel Medical), and the results were compared. Intraocular lens (IOL) power calculations in the same eyes after cataract extraction and posterior chamber IOL implantation were evaluated retrospectively based on the postoperative spherical equivalent prediction error.
RESULTS: Immersion ultrasonography and partial coherence interferometry measurements correlated in a highly positive manner (correlation coefficient = 0.996). Outcomes analysis demonstrated 92.0% of eyes were within +/-0.5 diopter of emmetropia based on immersion axial length measurements.
CONCLUSION: Immersion ultrasonography provided highly accurate axial length measurements and permitted highly accurate IOL power calculations.
Full text links
Trending Papers
Management of Latent Tuberculosis Infection.JAMA 2023 January 20
The Difficult Airway Redefined.Prehospital and Disaster Medicine 2022 November 10
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app