Effect of acetabular modularity on polyethylene wear and osteolysis in total hip arthroplasty.
Journal of Bone and Joint Surgery. American Volume 2002 January
BACKGROUND: Debris from polyethylene wear causes osteolysis. In this study, we examined the effect of acetabular liner modularity on polyethylene wear and osteolysis.
METHODS: We compared forty-one hips (thirty-nine patients) treated with a nonmodular, porous-coated acetabular component with a matched group of forty-one hips (forty patients) treated with a modular acetabular component. The groups were matched by patient gender and age, type of polyethylene material, method of polyethylene sterilization, femoral head size and manufacturer, and stem manufacturer. The mean follow-up period was 5.3 years (range, 3.8 to 6.8 years) for the nonmodular group and 5.5 years (range, 3.8 to 8.0 years) for the modular group. Using serial radiographs and a computer-assisted method, we measured two-dimensional head penetration into the polyethylene liner. Temporal head-penetration data and linear regression analysis were used to calculate the true wear rates.
RESULTS: The nonmodular acetabular components demonstrated a lower, but not a significantly lower, mean true wear rate than did the modular components (0.11 compared with 0.16 mm/yr, p = 0.22), and they were associated with a significantly lower rate of osteolysis (2% compared with 22%, p = 0.01). In addition, the true wear rates of the nonmodular components were less variable than those of the modular components. The 95% confidence interval for the wear rates of the nonmodular components (0.08 to 0.13 mm/yr) was nearly half that of the modular group (0.11 to 0.20 mm/yr).
CONCLUSIONS: The lower and more consistent true wear rates of the nonmodular components could be attributed to the fact that these cups were designed to have greater liner-shell conformity, greater liner thickness, and less liner-shell micromotion than modular components. These design factors could have favorably altered the stress distribution throughout the liner and could have thereby decreased wear. Although nonmodular components may present a partial solution to the problems of wear and osteolysis, they pose a disadvantage when a failed liner in a bone-ingrown acetabular component needs to be revised.
METHODS: We compared forty-one hips (thirty-nine patients) treated with a nonmodular, porous-coated acetabular component with a matched group of forty-one hips (forty patients) treated with a modular acetabular component. The groups were matched by patient gender and age, type of polyethylene material, method of polyethylene sterilization, femoral head size and manufacturer, and stem manufacturer. The mean follow-up period was 5.3 years (range, 3.8 to 6.8 years) for the nonmodular group and 5.5 years (range, 3.8 to 8.0 years) for the modular group. Using serial radiographs and a computer-assisted method, we measured two-dimensional head penetration into the polyethylene liner. Temporal head-penetration data and linear regression analysis were used to calculate the true wear rates.
RESULTS: The nonmodular acetabular components demonstrated a lower, but not a significantly lower, mean true wear rate than did the modular components (0.11 compared with 0.16 mm/yr, p = 0.22), and they were associated with a significantly lower rate of osteolysis (2% compared with 22%, p = 0.01). In addition, the true wear rates of the nonmodular components were less variable than those of the modular components. The 95% confidence interval for the wear rates of the nonmodular components (0.08 to 0.13 mm/yr) was nearly half that of the modular group (0.11 to 0.20 mm/yr).
CONCLUSIONS: The lower and more consistent true wear rates of the nonmodular components could be attributed to the fact that these cups were designed to have greater liner-shell conformity, greater liner thickness, and less liner-shell micromotion than modular components. These design factors could have favorably altered the stress distribution throughout the liner and could have thereby decreased wear. Although nonmodular components may present a partial solution to the problems of wear and osteolysis, they pose a disadvantage when a failed liner in a bone-ingrown acetabular component needs to be revised.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app