Daily targeting of intrahepatic tumors for radiotherapy.

INTRODUCTION: A system has been developed for daily targeting of intrahepatic tumors using a combination of ventilatory immobilization, in-room diagnostic imaging, and on-line setup adjustment. By reducing geometric position uncertainty, as well as organ movement, this system permits reduction of margins and thus potentially higher treatment doses. This paper reports our initial experience treating 8 patients with focal liver tumors using this system.

METHODS AND MATERIALS: The system includes diagnostic X-ray tubes mounted on the wall and ceiling of a treatment room, an active matrix flat panel imager, in-room control for image acquisition and setup adjustment, and a ventilatory immobilization system via active breathing control (ABC). Eight patients participated in the study, two using an early prototype ABC unit, and the remaining six with a commercial ABC system and improved setup measurement tools. Treatment margins were reduced, and dose consequently increased because of increased confidence in target position under this protocol. After daily setup via skin marks, orthogonal radiographs were acquired at suspended ventilation. The images were aligned to the CT model using the diaphragm for inferior-superior (IS) alignment, and the skeleton for left-right (LR) and anterior-posterior (AP) alignment. Adjustments were made for positioning errors greater than a threshold (3 or 5 mm). After treatment, retrospective analysis determined the final setup accuracy, as well as the error in initial setup measurement performed before setup adjustment.

RESULTS: Two hundred sixty-two treatment fractions were delivered on eight patients, with 171 treatments requiring repositioning. Typical treatment times were 25-30 min. Patients were able to tolerate ABC throughout the course of treatment. Breath holds up to 35 s long were used for treatment. The use of on-line imaging and setup adjustment reduced setup errors (sigma) from 4.0 mm (LR), 6.7 mm (IS), and 3.8 mm (AP) to 2.1 mm (LR), 3.5 mm (IS), and 2.3 mm (AP). Prescribed doses were increased using this system by an average of 5 Gy.

CONCLUSIONS: Daily targeting of intrahepatic targets has been demonstrated to be feasible. The potential for reduction in treatment margin and consequential safe dose escalation has been demonstrated, while maintaining reasonable treatment delivery times.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app