Value of high-resolution MR in patients scheduled for cochlear implantation.
Acta Radiologica 2001 November
OBJECTIVE: To determine sensitivity and specificity of high-resolution MR imaging and of high-resolution axial CT (HRCT) and to compare the two modalities in predicting the surgical and functional success of cochlear implantation.
MATERIAL AND METHODS: The presurgical MR images (2D T2W TSE, 3D T2*W CISS, plain and contrast-enhanced 3D T1W MP-RAGE) and axial HRCT findings of 26 patients were evaluated with regard to the predictive value concerning the success of cochlear implantation.
RESULTS: We found a high correlation between MR and HRCT and the success of cochlear implantation. In all 26 patients, the MR-based predictions concerning the success of cochlear implantation were correct. In 10 patients, MR gave additional information to HRCT. In all patients, MR gave sufficient information about the status of the inner ear, inner auditory canal and cochlear nerve to aid the surgeon during the operation.
CONCLUSION: A high-resolution MR protocol consisting of coronal 2D T2W TSE, 3D T2*W axial CISS, plain and contrast-enhanced sagittal T1W 3D MP-RAGE is recommended for the evaluation of candidates scheduled for cochlear implantation. It provides information which cannot be obtained by HRCT.
MATERIAL AND METHODS: The presurgical MR images (2D T2W TSE, 3D T2*W CISS, plain and contrast-enhanced 3D T1W MP-RAGE) and axial HRCT findings of 26 patients were evaluated with regard to the predictive value concerning the success of cochlear implantation.
RESULTS: We found a high correlation between MR and HRCT and the success of cochlear implantation. In all 26 patients, the MR-based predictions concerning the success of cochlear implantation were correct. In 10 patients, MR gave additional information to HRCT. In all patients, MR gave sufficient information about the status of the inner ear, inner auditory canal and cochlear nerve to aid the surgeon during the operation.
CONCLUSION: A high-resolution MR protocol consisting of coronal 2D T2W TSE, 3D T2*W axial CISS, plain and contrast-enhanced sagittal T1W 3D MP-RAGE is recommended for the evaluation of candidates scheduled for cochlear implantation. It provides information which cannot be obtained by HRCT.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app