JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Solid phase assay of urine cystine supersaturation in the presence of cystine binding drugs.
Journal of Urology 2001 August
PURPOSE: We report a new kind of assay system for urine cystine supersaturation that is accurate in the presence of cystine binding thiol drugs. We measured the molar ratio of cystine dissolved per mole of drug.
MATERIALS AND METHODS: Measured amounts of cystine crystals were incubated in buffer or urine for 48 hours with stirring. The solid phase remaining was pelleted by centrifugation, extracted into a high pH buffer and measured. D-penicillamine, tiopronin and captopril were added to determine their effect on solid phase dissolution.
RESULTS: Total cystine calculated from urine and solid measurements closely matched the amounts of cystine weighed in, meaning that the assay system successfully recovered the total cystine from the 2 phases. Each drug dissolved solid cystine in a specific and fixed proportion to its molar concentration in the range of 0.2 to 0.4 mM. dissolution per mM. of drug. Solution measurements were not a reliable gauge to the actual amounts of cystine dissolved.
CONCLUSIONS: Changes in solid phase cystine accurately reflect buffer or urine supersaturation when thiol drugs are present. The solid phase assay is a technically straightforward and reliable way of assessing cystine movement into and out of urine that avoids complexity of measurement and distortions of assay systems by drugs. This assay enables one to assess the level of drug effect and the need for a change in dosing.
MATERIALS AND METHODS: Measured amounts of cystine crystals were incubated in buffer or urine for 48 hours with stirring. The solid phase remaining was pelleted by centrifugation, extracted into a high pH buffer and measured. D-penicillamine, tiopronin and captopril were added to determine their effect on solid phase dissolution.
RESULTS: Total cystine calculated from urine and solid measurements closely matched the amounts of cystine weighed in, meaning that the assay system successfully recovered the total cystine from the 2 phases. Each drug dissolved solid cystine in a specific and fixed proportion to its molar concentration in the range of 0.2 to 0.4 mM. dissolution per mM. of drug. Solution measurements were not a reliable gauge to the actual amounts of cystine dissolved.
CONCLUSIONS: Changes in solid phase cystine accurately reflect buffer or urine supersaturation when thiol drugs are present. The solid phase assay is a technically straightforward and reliable way of assessing cystine movement into and out of urine that avoids complexity of measurement and distortions of assay systems by drugs. This assay enables one to assess the level of drug effect and the need for a change in dosing.
Full text links
Trending Papers
Fluid Resuscitation in Patients With Traumatic Brain Injury: A Comprehensive Review.Curēus 2023 August
Glycaemic Control and Weight Reduction: A Narrative Review of New Therapies for Type 2 Diabetes.Diabetes Therapy : Research, Treatment and Education of Diabetes and related Disorders 2023 September 16
Differential Diagnosis and Therapeutic Advances in Multiple Myeloma: A Review Article.Blood and Lymphatic Cancer : Targets and Therapy 2023
Intraoperative use of phenylephrine versus ephedrine and postoperative delirium: A multicenter retrospective cohort study.Anesthesiology 2023 September 20
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app