[Usefulness of a VDD defibrillation electrode in recording atrial electrograms during atrial flutter and atrial fibrillation].

Herz 2001 Februrary
BACKGROUND: Monitoring of atrial signals improves the accuracy in identifying supraventricular tachyarrhythmias to prevent inappropriate therapies in patients with implantable cardioverter-defibrillators (ICD). Since complications due to the additional atrial lead were found in dual chamber ICD systems with 2 leads, we designed a single-pass VDD-lead for use with dual chamber ICDs.

PATIENTS AND METHODS: After promising animal experiments in a German multicenter study a prototype VDD lead (single-coil defibrillation electrode with 2 additional fractally coated rings for bipolar sensing in the atrium) was temporarily used in 20 patients. Atrial and ventricular signals were recorded during sinus rhythm, atrial flutter, atrial fibrillation and ventricular tachycardia or ventricular fibrillation. Terminations of ventricular arrhythmias were performed by internal DC shock.

RESULTS: The implantation of the electrode was successful in 18 of 20 patients. Mean atrial pacing threshold was 2.45 +/- 0.9 V/0.5 ms, mean atrial impedance was 215 +/- 31 Ohm. Atrial amplitudes were greater during sinus rhythm (2.7 +/- 1.6 mV) than during atrial flutter (1.36 +/- 0.28 mV, p < 0.05) or atrial fibrillation (0.92 +/- 0.29 mV, p < 0.01). During ventricular fibrillation atrial "sinus"-signals had significantly (p < 0.01) lower amplitudes than during sinus rhythm. Mean ventricular sensing was 13.3 +/- 7.9 mV, mean ventricular impedance was 577 +/- 64 Ohm. Defibrillation was successful with 20 J shock. 99.6% of P waves could be detected in sinus rhythm and 85 +/- 9.9% of flutter waves during atrial flutter. During atrial fibrillation 55% of atrial signals could be detected without modification of the signal amplifier.

CONCLUSIONS: A new designed VDD dual chamber electrode provides stable detection of atrial and ventricular signals during sinus rhythm and atrial flutter. For reliable detection of atrial fibrillation modifications of the signal amplifier are necessary.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app