Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Antiarrhythmics vs Implantable Defibrillator study. Cardiac Arrest Study Hamburg . Canadian Implantable Defibrillator Study.

AIMS: Three randomized trials of implantable cardioverter defibrillator (ICD) therapy vs medical treatment for the prevention of death in survivors of ventricular fibrillation or sustained ventricular tachycardia have been reported with what might appear to be different results. The present analysis was performed to obtain the most precise estimate of the efficacy of the ICD, compared to amiodarone, for prolonging survival in patients with malignant ventricular arrhythmia.

METHODS AND RESULTS: Individual patient data from the Antiarrhythmics vs Implantable Defibrillator (AVID) study, the Cardiac Arrest Study Hamburg (CASH) and the Canadian Implantable Defibrillator Study (CIDS) were merged into a master database according to a pre-specified protocol. Proportional hazard modelling of individual patient data was used to estimate hazard ratios and to investigate subgroup interactions. Fixed effect meta-analysis techniques were also used to evaluate treatment effects and to assess heterogeneity across studies. The classic fixed effects meta-analysis showed that the estimates of ICD benefit from the three studies were consistent with each other (P heterogeneity=0.306). It also showed a significant reduction in death from any cause with the ICD; with a summary hazard ratio (ICD:amiodarone) of 0.72 (95% confidence interval 0.60, 0.87;P=0.0006). For the outcome of arrhythmic death, the hazard ratio was 0.50 (95% confidence interval 0.37, 0.67;P<0.0001). Survival was extended by a mean of 4.4 months by the ICD over a follow-up period of 6 years. Patients with left ventricular ejection fraction < or = 35% derived significantly more benefit from ICD therapy than those with better preserved left ventricular function. Patients treated before the availability of non-thoracotomy ICD implants derived significantly less benefit from ICD therapy than those treated in the non-thoracotomy era.

CONCLUSION: Results from the three trials of the ICD vs amiodarone are consistent with each other. There is a 28% reduction in the relative risk of death with the ICD that is due almost entirely to a 50% reduction in arrhythmic death.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app