Use of automated external defibrillators by a U.S. airline.
New England Journal of Medicine 2000 October 27
BACKGROUND: Passengers who have ventricular fibrillation aboard commercial aircraft rarely survive, owing to the delay in obtaining emergency care and defibrillation.
METHODS: In 1997, a major U.S. airline began equipping its aircraft with automated external defibrillators. Flight attendants were trained in the use of the defibrillator and applied the device when passengers had a lack of consciousness, pulse, or respiration. The automated external defibrillator was also used as a monitor for other medical emergencies, generally at the direction of a passenger who was a physician. The electrocardiogram that was obtained during each use of the device was analyzed by two arrhythmia specialists for appropriateness of use. We analyzed data on all 200 instances in which the defibrillators were used between June 1, 1997, and July 15, 1999.
RESULTS: Automated external defibrillators were used for 200 patients (191 on the aircraft and 9 in the terminal), including 99 with documented loss of consciousness. Electrocardiographic data were available for 185 patients. The administration of shock was advised in all 14 patients who had electrocardiographically documented ventricular fibrillation, and no shock was advised in the remaining patients (sensitivity and specificity of the defibrillator in identifying ventricular fibrillation, 100 percent). The first shock successfully defibrillated the heart in 13 patients (defibrillation was withheld in 1 case at the family's request). The rate of survival to discharge from the hospital after shock with the automated external defibrillator was 40 percent. A total of 36 patients either died or were resuscitated after cardiac arrest. No complications arose from use of the automated external defibrillator as a monitor in conscious passengers.
CONCLUSIONS: The use of the automated external defibrillator aboard commercial aircraft is effective, with an excellent rate of survival to discharge from the hospital after conversion of ventricular fibrillation. There are not likely to be complications when the device is used as a monitor in the absence of ventricular fibrillation.
METHODS: In 1997, a major U.S. airline began equipping its aircraft with automated external defibrillators. Flight attendants were trained in the use of the defibrillator and applied the device when passengers had a lack of consciousness, pulse, or respiration. The automated external defibrillator was also used as a monitor for other medical emergencies, generally at the direction of a passenger who was a physician. The electrocardiogram that was obtained during each use of the device was analyzed by two arrhythmia specialists for appropriateness of use. We analyzed data on all 200 instances in which the defibrillators were used between June 1, 1997, and July 15, 1999.
RESULTS: Automated external defibrillators were used for 200 patients (191 on the aircraft and 9 in the terminal), including 99 with documented loss of consciousness. Electrocardiographic data were available for 185 patients. The administration of shock was advised in all 14 patients who had electrocardiographically documented ventricular fibrillation, and no shock was advised in the remaining patients (sensitivity and specificity of the defibrillator in identifying ventricular fibrillation, 100 percent). The first shock successfully defibrillated the heart in 13 patients (defibrillation was withheld in 1 case at the family's request). The rate of survival to discharge from the hospital after shock with the automated external defibrillator was 40 percent. A total of 36 patients either died or were resuscitated after cardiac arrest. No complications arose from use of the automated external defibrillator as a monitor in conscious passengers.
CONCLUSIONS: The use of the automated external defibrillator aboard commercial aircraft is effective, with an excellent rate of survival to discharge from the hospital after conversion of ventricular fibrillation. There are not likely to be complications when the device is used as a monitor in the absence of ventricular fibrillation.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app