We have located links that may give you full text access.
Journal Article
Research Support, Non-U.S. Gov't
Intersample fluctuations in phosphocreatine concentration determined by 31P-magnetic resonance spectroscopy and parameter estimation of metabolic responses to exercise in humans.
Journal of Physiology 2000 October 16
The ATP turnover rate during constant-load exercise is often estimated from the initial rate of change of phosphocreatine concentration ([PCr]) using 31P-magnetic resonance spectroscopy (MRS). However, the phase and amplitude characteristics of the sample-to-sample fluctuations can markedly influence this estimation (as well as that for the time constant (tau) of the [PCr] change) and confound its physiological interpretation especially for small amplitude responses. This influence was investigated in six healthy males who performed repeated constant-load quadriceps exercise of a moderate intensity in a whole-body MRS system. A transmit- receive surface coil was placed under the right quadriceps, allowing determination of intramuscular [PCr]; pulmonary oxygen uptake (VO2) was simultaneously determined, breath-by-breath, using a mass spectrometer and a turbine volume measuring module. The probability density functions (PDF) of [PCr] and VO2 fluctuations were determined for each test during the steady states of rest and exercise and the PDF was then fitted to a Gaussian function. The standard deviation of the [PCr] and VO2 fluctuations at rest and during exercise (sr and sw, respectively) and the peak centres of the distributions (xc(r) and xc(w)) were determined, as were the skewness (gamma1) and kurtosis (gamma2) coefficients. There was no difference between sr and sw for [PCr] relative to the resting control baseline (s(r) = 1.554 %delta (s.d. = 0.44), s(w) = 1.514 %delta (s.d. = 0.35)) or the PDF peak centres (xc(r) = -0.013 %delta (s.d. = 0.09), xc(w) -0.197 %delta (s.d. = 0.18)). The standard deviation and peak centre of the 'noise' in VO2 also did not vary between rest and exercise (sr = 0.0427 l min(-1) (s.d. = 0.0104), s(w) = 0.0640 l min(-1) (s.d. = 0.0292); xc(r) = -0.0051 l min(-1) (s.d. = 0.0069), xc(w) 0.0022 l min(-1) (s.d. = 0.0034)). Our results demonstrate that the intersample 'noise' associated with [PCr] determination by 31P-MRS may be characterised as a stochastic Gaussian process that is uncorrelated with work rate, as previously described for VO2. This 'noise' can significantly affect the estimation of tau[PCr] and especially the initial rate of change of [PCr], i.e. the fluctuations can lead to variations in estimation of the initial rate of change of [PCr] of more than twofold, if the inherent 'noise' is not accounted for. This 'error' may be significantly reduced in such cases if the initial rate of change is estimated from the time constant and amplitude of the response.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app