OBJECTIVE: Comparison of suprasternal Doppler (SST) and thermodilution (TD) for the measurement of cardiac output (CO) in critically ill patients.

DESIGN: Prospective study.

SETTING: Intensive care unit of a university hospital.

PATIENTS AND PARTICIPANTS: 65 consecutive critically ill patients requiring a pulmonary artery catheter.

INTERVENTIONS: Paired CO measurements were made simultaneously using SST and TD by two independent operators. The time to obtain a CO value by SST was measured. Correlation coefficients and the linear regression equation were determined. A Bland and Altman diagram was plotted. A Bland and Altman diagram was also plotted for the level of cardiac index (CI) values (low: CI < 2.5 l min(-1) m(-2); normal: 2.5 < or = CI < or = 4.5 l min(-1) m(-2); high: CI > 4.5 l min(-1) m(-2)).

MEASUREMENTS AND RESULTS: In seven patients SST failed to measure CO. In the remaining 58 patients 314 paired CO measurements were performed. The mean time to measure CO by SST was 73 +/- 45 s. The equation of linear regression was: SST(CO) = 0.84 TD(CO) + 1.39. The correlation coefficient was 0.84. The bias between SST and TD was -0.2 +/- 1.4 l min(-1). Biases were -0.23 +/- 0.50, -0.20 +/- 0.68, and 0.25 +/- 0.92 l min(-1) m(-2) for low, normal, and high levels of CI, respectively.

CONCLUSION: SST does not accurately measure CO but allows a rapid assessment of CI level in critically ill patients.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Remove bar
Read by QxMD icon Read
10945385
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"