Assessment of cardiac preload and extravascular lung water by single transpulmonary thermodilution

S G Sakka, C C Rühl, U J Pfeiffer, R Beale, A McLuckie, K Reinhart, A Meier-Hellmann
Intensive Care Medicine 2000, 26 (2): 180-7
10784306

OBJECTIVE: Transpulmonary double-indicator dilution is a useful monitoring technique for measurement of intrathoracic blood volume (ITBV) and extravascular lung water (EVLW). In this study, we compared a simpler approach using single arterial thermodilution derived measurements of ITBV and EVLW with the double-indicator dilution technique.

DESIGN: Prospective observational clinical study.

SETTING: Surgical intensive care units of two university hospitals.

PATIENTS AND METHODS: Global end-diastolic volume (GEDV) derived from single thermodilution was used for calculation of ITBV. Structural regression analysis of the first two thermo-dye dilution measurements in a derivation population of 57 critically ill patients (38 male, 19 female, 18-79 years, 56 +/- 15 years) revealed ITBV = (1.25 x GEDV)-28.4 (ml). This equation was then applied to all first measurements in a validation population of 209 critically ill patients (139 male, 70 female, 10-88 years, mean 53 +/- 19 years), and single-thermodilution ITBV (ITBV(ST)) and EVLW (EVLW(ST)) was calculated and compared to thermo-dye dilution derived values (ITBV(TD), EVLW(TD)). For inter-individual comparison, absolute values for ITBV and EVLW were normalised as indexed by body surface area (ITBVI) and body weight (EVLWI), respectively.

MEASUREMENTS AND RESULTS: Linear regression analysis yielded a correlation of ITBVI(ST) = (1.05 x ITBVI(TD))-58.0 (ml/m2), r = 0.97, P < 0.0001. Bias between ITBVI(TD) and ITBVI(ST) was 7.6 (ml/m2) with a standard deviation of 57.4 (ml/m2). Single-thermodilution EVLWI (EVLWI(ST)) was calculated using ITBVI(ST) and revealed the correlation EVLWI(ST) = (0.83 x EVLWI(TD)) + 1.6 (ml/kg), r = 0.96, P < 0.0001. Bias between EVLWI(TD) and EVLWI(ST) was -0.2 (ml/kg) with a standard deviation of 1.4 (ml/kg). In detail, EVLWI(ST) systematically overestimated EVLWI(TD) at low-normal values for EVLWI and underestimated EVLWI at higher values (above 12 ml/kg).

CONCLUSION: Determinations of ITBV and EVLW by single thermodilution agreed closely with the corresponding values from the double-indicator technique. Since transpulmonary single thermodilution is simple to apply, less invasive and cheaper, all these features make it a promising technique for the bedside. Nevertheless, further validation studies are needed in the future.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Trending Papers

Remove bar
Read by QxMD icon Read
10784306
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"