PURPOSE: Tunneled catheters are an alternative means of vascular access for patients in need of hemodialysis who cannot undergo dialysis through a surgical shunt. This study was undertaken to evaluate the performance of the Tesio dialysis catheter.
MATERIALS AND METHODS: A prospective study of the Tesio catheter was performed. Follow-up data regarding catheter function and adequacy of dialysis were obtained from nine hemodialysis facilities.
RESULTS: Seventy-nine Tesio catheters were placed in 71 patients. Immediate technical success was 99% (78 of 79 catheters). The procedure complication rate was 9% (seven catheters). Only two complications required intervention: one fatal air embolism and one chest wall hematoma. Sixty-seven catheters in 60 patients were followed up for a total of 4,367 catheter days. Overall, catheter-related infection occurred in 9% (six of 67 catheters). Primary catheter patency was 87% at 1 week, 82% at 1 month, 72% at 3 months, and 66% at 6 months. Mean blood flow was 286 mL/min immediately after insertion, 301 mL/min at 3 months, and 306 mL/min at 6 months. Adequate dialysis dose as reflected by a urea reduction ratio of 60 or more or a urea kinetic modeling, or Kt/V, value of 1.2 or more was observed on at least one occasion for 74% and 76% of catheters, respectively.
CONCLUSION: The Tesio catheter is a reasonable means of vascular access for patients who undergo dialysis but are not candidates for surgical shunt placement.
MATERIALS AND METHODS: A prospective study of the Tesio catheter was performed. Follow-up data regarding catheter function and adequacy of dialysis were obtained from nine hemodialysis facilities.
RESULTS: Seventy-nine Tesio catheters were placed in 71 patients. Immediate technical success was 99% (78 of 79 catheters). The procedure complication rate was 9% (seven catheters). Only two complications required intervention: one fatal air embolism and one chest wall hematoma. Sixty-seven catheters in 60 patients were followed up for a total of 4,367 catheter days. Overall, catheter-related infection occurred in 9% (six of 67 catheters). Primary catheter patency was 87% at 1 week, 82% at 1 month, 72% at 3 months, and 66% at 6 months. Mean blood flow was 286 mL/min immediately after insertion, 301 mL/min at 3 months, and 306 mL/min at 6 months. Adequate dialysis dose as reflected by a urea reduction ratio of 60 or more or a urea kinetic modeling, or Kt/V, value of 1.2 or more was observed on at least one occasion for 74% and 76% of catheters, respectively.
CONCLUSION: The Tesio catheter is a reasonable means of vascular access for patients who undergo dialysis but are not candidates for surgical shunt placement.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app