Automated analysis of digital oximetry in the diagnosis of obstructive sleep apnoea.
Thorax 2000 April
BACKGROUND: The gold standard diagnostic test for obstructive sleep apnoea (OSA) is overnight polysomnography (PSG) which is costly in terms of time and money. Consequently, a number of alternatives to PSG have been proposed. Oximetry is appealing because of its widespread availability and ease of application. The diagnostic performance of an automated analysis algorithm based on falls and recovery of digitally recorded oxygen saturation was compared with PSG.
METHODS: Two hundred and forty six patients with suspected OSA were randomly selected for PSG and automated off line analysis of the digitally recorded oximeter signal.
RESULTS: The PSG derived apnoea hypopnea index (AHI) and oximeter derived respiratory disturbance index (RDI) were highly correlated (R = 0.97). The mean (2SD) of the differences between AHI and RDI was 2.18 (12.34)/h. The sensitivity and specificity of the algorithm depended on the AHI and RDI criteria selected for OSA case designation. Using case designation criteria of 15/h for AHI and RDI, the sensitivity and specificity were 98% and 88%, respectively. If the PSG derived AHI included EEG based arousals as part of the hypopnea definition, the mean (2SD) of the differences between RDI and AHI was -0.12 (15. 62)/h and the sensitivity and specificity profile did not change significantly.
CONCLUSIONS: In a population of patients suspected of having OSA, off line automated analysis of the oximetry signal provides a close estimate of AHI as well as excellent diagnostic sensitivity and specificity for OSA.
METHODS: Two hundred and forty six patients with suspected OSA were randomly selected for PSG and automated off line analysis of the digitally recorded oximeter signal.
RESULTS: The PSG derived apnoea hypopnea index (AHI) and oximeter derived respiratory disturbance index (RDI) were highly correlated (R = 0.97). The mean (2SD) of the differences between AHI and RDI was 2.18 (12.34)/h. The sensitivity and specificity of the algorithm depended on the AHI and RDI criteria selected for OSA case designation. Using case designation criteria of 15/h for AHI and RDI, the sensitivity and specificity were 98% and 88%, respectively. If the PSG derived AHI included EEG based arousals as part of the hypopnea definition, the mean (2SD) of the differences between RDI and AHI was -0.12 (15. 62)/h and the sensitivity and specificity profile did not change significantly.
CONCLUSIONS: In a population of patients suspected of having OSA, off line automated analysis of the oximetry signal provides a close estimate of AHI as well as excellent diagnostic sensitivity and specificity for OSA.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app