Patient radiation dose associated with uterine artery embolization.
Radiology 2000 January
PURPOSE: To evaluate the estimated absorbed radiation doses to the ovaries and skin entrance during uterine artery embolization (UAE) for leiomyomas.
MATERIALS AND METHODS: Radiation dose was measured in 20 patients who underwent UAE for leiomyomas. Measurements were obtained by placing lithium fluoride dosimeters both into the posterior fornix of the vagina and on the skin at the beam entrance site. Patient doses were obtained with thermoluminescent dosimeters.
RESULTS: The mean fluoroscopic time was 21.89 minutes, and the mean number of angiographic exposures was 44. The mean estimated absorbed ovarian dose was 22.34 cGy, and the mean absorbed skin dose was 162.32 cGy. These values compare to published values for the assessed absorbed ovarian dose during hysterosalpingography (0.04-0.55 cGy), fallopian tube recanalization (0.2-2.75 cGy), computed tomography of the trunk (0.1-1.9 cGy), and pelvic irradiation for Hodgkin disease (263-3,500 cGy).
CONCLUSION: The estimated absorbed ovarian dose during UAE is greater than that during common fluoroscopic procedures. On the basis of the known risks of pelvic irradiation for Hodgkin disease, the dose associated with UAE is unlikely to result in acute or long-term radiation injury to the patient or to a measurable increase in the genetic risk to the patient's future children.
MATERIALS AND METHODS: Radiation dose was measured in 20 patients who underwent UAE for leiomyomas. Measurements were obtained by placing lithium fluoride dosimeters both into the posterior fornix of the vagina and on the skin at the beam entrance site. Patient doses were obtained with thermoluminescent dosimeters.
RESULTS: The mean fluoroscopic time was 21.89 minutes, and the mean number of angiographic exposures was 44. The mean estimated absorbed ovarian dose was 22.34 cGy, and the mean absorbed skin dose was 162.32 cGy. These values compare to published values for the assessed absorbed ovarian dose during hysterosalpingography (0.04-0.55 cGy), fallopian tube recanalization (0.2-2.75 cGy), computed tomography of the trunk (0.1-1.9 cGy), and pelvic irradiation for Hodgkin disease (263-3,500 cGy).
CONCLUSION: The estimated absorbed ovarian dose during UAE is greater than that during common fluoroscopic procedures. On the basis of the known risks of pelvic irradiation for Hodgkin disease, the dose associated with UAE is unlikely to result in acute or long-term radiation injury to the patient or to a measurable increase in the genetic risk to the patient's future children.
Full text links
Trending Papers
Carvedilol, probably the β-blocker of choice for everyone with cirrhosis and portal hypertension: But not so fast!Liver International : Official Journal of the International Association for the Study of the Liver 2023 June
Evidence-Based Guideline for the diagnosis and management of eosinophilic granulomatosis with polyangiitis.Nature Reviews. Rheumatology 2023 May 10
Advances in Acute Ischemic Stroke Treatment: Current Status and Future Directions.AJNR. American Journal of Neuroradiology 2023 May 19
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app