CLINICAL TRIAL
JOURNAL ARTICLE
RANDOMIZED CONTROLLED TRIAL
Add like
Add dislike
Add to saved papers

Patient-ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger.

OBJECTIVE: Air leaks around the mask are very likely to occur during noninvasive ventilation, in particular when prolonged ventilatory treatment is required. It has been suggested that leaks from the mask may impair the expiratory trigger cycling mechanism when inspiratory pressure support ventilation (PSV) is used. The aim of this study was to compare the short-term effect of two different expiratory cycling mechanisms (time-cycled vs flow-cycled) during noninvasive inspiratory pressure support ventilation (NIPSV) on patient-ventilator synchronisation in severe hypoxemic respiratory failure.

STUDY POPULATION: Six patients with acute lung injury (ALI) due to acquired immunodeficiency syndrome (AIDS)-related opportunistic pneumonia were enrolled in the protocol.

INTERVENTION: Each subject was first studied during spontaneous breathing with a Venturi oxygen mask (SB) and successively submitted to a randomly assigned 20' conventional flow-cycling (NIPSVfc) or time-cycling inspiratory pressure support ventilation (NIPSVtc). The pre-set parameters were: inspiratory pressure of 10 cm H2O, PEEP of 5 cm H2O for the same inspired oxygen fraction as during SB. A tight fit of the mask was avoided in order to facilitate air leaks around the mask. The esophageal pressure time product (PTPes) and tidal swings (delta Pes) were measured to evaluate the patient's respiratory effort. A subjective "comfort score" and the difference between patient and machine respiratory rate [delta RR(p-v)], calculated on esophageal and airway pressure curves, were used as indices of patient-machine interaction.

RESULTS: Air leaks through the mask occurred in five out of six patients. The values of PEEPi (< 1.9 cm H2O) excluded significant expiratory muscle activity. NIPSVtc significantly reduced PTPes, delta Pes, and delta RR(p-v) when compared to NIPS-Vfc [230 +/- 41 (SE) vs 376 +/- 72 cm H2O.s.min-1; 8 +/- 2 vs 13 +/- 2 cm H2O; 1 +/- 1 vs 9 +/- 2 br.min-1; respectively] with a concomitant significant improvement of the "comfort score".

CONCLUSIONS: In the presence of air leaks a time-cycled expiratory trigger provides a better patient-machine interaction than a flow-cycled expiratory trigger during NIPSV.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app