The effects of surface roughness and polymethylmethacrylate precoating on the radiographic and clinical results of the Iowa hip prosthesis. A study of patients less than fifty years old.
BACKGROUND: The purpose of the present study was to compare the results associated with two different surface finishes (bead-blasted and grit-blasted) for Iowa femoral components used in total hip arthroplasties performed between January 1979 and June 1991 in patients who were less than fifty years old.
METHODS: Between January 1979 and December 1985, thirty-six primary total hip replacements were performed in twenty-five patients with insertion of a bead-blasted Iowa femoral component (average surface roughness, 0.8 micrometer) with cement and insertion of a titanium-backed acetabular component (thirty-five hips) or non-metal-backed acetabular component (one hip) with cement. Between January 1986 and June 1991, forty-five primary total hip replacements were performed in thirty-seven patients with use of a precoated grit-blasted Iowa femoral component (average surface roughness, 2.1 micrometers) and a Harris-Galante-I porous ingrowth acetabular component (forty-one hips) or an Osteonics component (four hips). The only change in the design of the femoral component (other than the surface finish) between the two consecutive series was the addition of polymethylmethacrylate precoating to the proximal third of all forty-five grit-blasted stems and modularity of the femoral head of the last eight grit-blasted stems.
RESULTS: No hip was lost to follow-up. The duration of radiographic follow-up of the hips treated with the bead-blasted component averaged 11.3 years (range, ten to sixteen years) postoperatively, whereas that of the hips treated with the precoated grit-blasted component averaged 8.2 years (range, five to eleven years). Two (6 percent) of the bead-blasted femoral components were revised because of aseptic loosening compared with eight (18 percent) of the precoated grit-blasted components. Four bead-blasted components (11 percent) were either radiographically loose or were revised because of loosening compared with eleven precoated grit-blasted components (24 percent). Kaplan-Meier survivorship curves evaluated with log-rank analysis revealed that the bead-blasted Iowa femoral components were revised because of aseptic loosening (p = 0.0184) and were radiographically loose (p = 0.0068) less often than the precoated grit-blasted Iowa femoral components.
CONCLUSIONS: The findings of the present study have led the senior ones of us to resume using femoral components with a polished surface and fixed with cement.
METHODS: Between January 1979 and December 1985, thirty-six primary total hip replacements were performed in twenty-five patients with insertion of a bead-blasted Iowa femoral component (average surface roughness, 0.8 micrometer) with cement and insertion of a titanium-backed acetabular component (thirty-five hips) or non-metal-backed acetabular component (one hip) with cement. Between January 1986 and June 1991, forty-five primary total hip replacements were performed in thirty-seven patients with use of a precoated grit-blasted Iowa femoral component (average surface roughness, 2.1 micrometers) and a Harris-Galante-I porous ingrowth acetabular component (forty-one hips) or an Osteonics component (four hips). The only change in the design of the femoral component (other than the surface finish) between the two consecutive series was the addition of polymethylmethacrylate precoating to the proximal third of all forty-five grit-blasted stems and modularity of the femoral head of the last eight grit-blasted stems.
RESULTS: No hip was lost to follow-up. The duration of radiographic follow-up of the hips treated with the bead-blasted component averaged 11.3 years (range, ten to sixteen years) postoperatively, whereas that of the hips treated with the precoated grit-blasted component averaged 8.2 years (range, five to eleven years). Two (6 percent) of the bead-blasted femoral components were revised because of aseptic loosening compared with eight (18 percent) of the precoated grit-blasted components. Four bead-blasted components (11 percent) were either radiographically loose or were revised because of loosening compared with eleven precoated grit-blasted components (24 percent). Kaplan-Meier survivorship curves evaluated with log-rank analysis revealed that the bead-blasted Iowa femoral components were revised because of aseptic loosening (p = 0.0184) and were radiographically loose (p = 0.0068) less often than the precoated grit-blasted Iowa femoral components.
CONCLUSIONS: The findings of the present study have led the senior ones of us to resume using femoral components with a polished surface and fixed with cement.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app