Total hip arthroplasty with insertion of the acetabular component without cement in hips with total congenital dislocation or marked congenital dysplasia.

BACKGROUND: End-stage osteoarthritis secondary to total congenital dislocation or severe congenital dysplasia of the hip (class B or C according to the system of Eftekhar or type 2 or 3 according to the system of Hartofilakidis et al.) in adults presents special problems with regard to reconstruction of the hip. The purpose of the present study was to assess the intermediate-term results associated with the use of a porous ingrowth acetabular component for the treatment of these difficult cases.

METHODS: We performed a prospective study of a consecutive series of twenty-one patients (twenty-four hips) who had had a primary total hip arthroplasty with use of a hemispherical acetabular component that was inserted without cement and fixed with screws. No patient was lost to follow-up. Three patients (four hips) died, of causes unrelated to the total hip arthroplasty, before a minimum duration of follow-up of five years. None of these patients had had revision. Of the remaining eighteen patients (twenty hips), fifteen were women and three were men. Ten hips had total dislocation, and ten had severe dysplasia.

RESULTS: After an average duration of follow-up of eighty-three months (range, sixty-four to 102 months), the average Harris hip score was 90 points (range, 68 to 97 points). No patient had revision, loosening, or migration of the acetabular component; pelvic osteolysis; or a continuous radiolucent line at the mesh-bone interface of the acetabular component. The average rate of polyethylene wear was 0.08 millimeter per year (range, zero to 0.21 millimeter per year).

CONCLUSIONS: The porous ingrowth acetabular component that was used in the present study functioned well at the time of the intermediate-term follow-up of this group of patients who had marked congenital dysplasia or total dislocation of the hip. The use of this component decreased the need for structural acetabular grafts. This component appears to perform as well as larger components of this design that have been assessed after similar durations of follow-up.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app