Ventilator Alarms in Intensive Care Units: Frequency, Duration, Priority, and Relationship to Ventilator Parameters.

Ventilator alarms have long been presumed to contribute substantially to the overall alarm burden in the intensive care unit. In a prospective observational study, we determined that each ventilator triggered an alarm cascade of up to 8 separate notifications once every 6 minutes. In 1 intensive care unit with different ventilator manufacturers, the distribution of high-priority alarms was manufacturer dependent with 8.6% of alarms from 1 type and 89.8% of alarms from another type of ventilator. Alarm limits were not a function of patient-specific ventilator settings.